ImageXpress Micro High Content Imaging System Options User Guide
72
5014166 D
15.
Snap an image.
The condenser annulus ring shows. See
. You might
need to adjust the exposure time.
Figure 4-21:
Image of condenser annulus ring
16.
In the MetaXpress software, click
Live Mode
in either the
Acquire
or
Plate Acquisition and Control
dialog.
17.
Click
Regions > Load Regions
to load in the saved regions.
18.
Using the condenser annulus centering knobs located on the
right top and left top surface of the condenser turret, center the
condenser annulus ring with the objective phase ring. See
.
ImageXpressOptionsUserGuide.book Page 72 Wednesday, September 9, 2015 3:54 PM