ImageXpress Micro High Content Imaging System Options User Guide
38
5014166 D
8.
If applicable, for the
Sample Plate
, type a value for
X
,
Y
, and
Z
offsets from the well center (in
μ
m) for the sample plate.
Type a positive Z offset to have the pipettor sample higher than
normal. Type a negative Z offset to have the pipettor sample
lower than normal.
9.
If applicable, select
Track Volume
, and then type a value for the
Initial Volume for Tracking (
μ
l)
for each plate.
10.
Do one of the following:
If you selected
Track Volume
, then select
Track Liquid
Surface,
and then go to
.
Optionally, select
Track Volume,
select
Track Liquid Surface,
clear
Track Volume,
and then go to
.
Otherwise, go to
.
11.
Click
System Properties
.
Figure 2-7:
Fluidics System Properties dialog
CAUTION!
An incorrect X, Y, or Z offset can cause the pipette tip
to crash into the sample plate.
Note:
The
Track Liquid Surface
feature is used in conjunction
with the
Wet Dispense
feature. If
Track Liquid Surface
is
selected, then wet dispensing is done near the liquid surface.
ImageXpressOptionsUserGuide.book Page 38 Wednesday, September 9, 2015 3:54 PM