3. Operating Instructions
Table 3.4 (Continued) MENU Button and Menus
Menu Selection
Purpose/Action
Vert.
Ref Off/On
Toggles VR (Vertical Reference) mode on/off.
Set Vert.
Ref.
Performs VR calibration using short/open
terminators. When setup is complete, a green
vertical line indicates start of VR measurements.
Vert.
Ref.
Center Imp.
Adjusts VR calibration to use impedance matching
adapters and baluns.
Vert.
Ref.
Start
Sets start of VR at the position of the active cursor.
Vert.
Ref.
End
Sets end of VR at the position of the active cursor.
Cable Segments
Add New Segment
Create a new segment boundary at the active cursor.
The selected segment will be split in two, each with
the same initial Vp.
Delete Segment
Delete the currently selected segment. The segment
will be merged with the segment to the left.
Delete All Segments
Delete all segments. The CT100B will return to
normal operation. A single Vp now applies to the
entire trace.
Prev Segment
Select the segment to the left of the currently
selected segment.
Next Segment
Select the segment to the right of the currently
selected segment.
Reset All Vp
Reset the Vp of all segments to 1.000.
Restore All Vp
Restore Vp of all segments to the previous values.
Extra Long Mode
Extra Long Mode allows the CT100B to obtain
traces out to 40,000 ft. (12,000 m) (assumes
Vp=0.66). Extra Long Mode is slower than other
cable lengths and some instrument features are not
available. When in Extra Long Mode, the Cable
Length menu option will indicate Extra Long Mode
On. Selecting the cable length menu item will switch
off Extra Long Mode.
Vert.
Units
Select vertical units for a time-domain trace from
reflection coefficient (m
ρ
), impedance (Ω) or voltage
standing wave ratio (VSWR). A few features of the
CT100B only work with the reflection coefficient
setting.
Envelope Plot
Captures transient cable and connector faults.
Envelope Plot Off/On
Enable/disable envelope plot mode. When turned on
the CT100B begins capturing trace data and
displaying envelope plot information.
22