B. Operating Performance Checks
B.4.3
Pulse Amplitude Check
1. Pulse amplitude measurement should be taken with a device that can accurately measure
frequency, pulse width, period, and amplitude.
2. Set the device input impedance to 50 ohms.
3. Connect one end of the 36 in. reference cable to a capable pulse measurement device and
the other end of the 36 in. reference cable to the CT100B front panel cable connector. A
trace should appear on the device display.
4. Switch the CT100B into the Short pulse setting from the
Main
menu
Cable Len
setting.
5. Measure the pulse amplitude coming off the front panel cable connector. The averaged
pulse amplitude into 50 ohms should be between 290 mV and 310 mV.
6. Repeat pulse amplitude comparison for Medium and Long settings.
B.4.4
Pulse Width Check
1. Pulse width measurement should be taken with a device that can accurately measure
frequency, pulse width, period, and amplitude.
2. Set the device input impedance to 50 ohms.
3. Connect one end of the 36 in. reference cable to a capable pulse measurement device and
the other end of the 36 in. reference cable to the CT100B front panel cable connector. A
trace should appear on the device display.
4. Switch the CT100B into the Short pulse setting from the
Main
menu
Cable Len
setting.
5. Measure the pulse width coming off the front panel cable connector. The average pulse
width should be between 1.000-2.000
µ
sec for Short, 5.000-6.000
µ
sec for Medium, and
20.00-21.00
µ
sec for Long.
6. Repeat pulse width check for Medium and Long settings.
B.4.5
Pulse Repetition Period Check
1. Pulse repetition rate measurement should be taken with a device that can accurately
measure frequency, pulse width, period, and amplitude.
2. Set the device input impedance to 50 ohms.
3. Connect one end of the 36 in. reference cable to a capable pulse measurement device and
the other end of the 36 in. reference cable to the CT100B front panel cable connector. A
trace should appear on the device display.
4. Switch the CT100B into the Short pulse setting from the
Main
menu
Cable Len
setting.
5. Measure the pulse repetition rate coming off the front panel cable connector. The average
pulse repetition rate should be 4.000
±
0.002
µ
sec for Short, 20.00
±
0.02
µ
sec for Medium,
and 80.00
±
0.02
µ
sec for Long.
6. Repeat pulse width check for Medium and Long settings.
B.4.6
Pulse Frequency Check
1. Pulse frequency measurement should be taken with a device that can accurately measure
frequency, pulse width, period, and amplitude.
2. Set the device input impedance to 50 ohms.
3. Connect one end of the 36 in. reference cable to a capable pulse measurement device and
the other end of the 36 in. reference cable to the CT100B front panel cable connector. A
trace should appear on the device display.
4. Switch the CT100B into the Short pulse setting from the
Main
menu
Cable Len
setting.
CT100B TDR Cable Analyzers Operator’s Manual
89