3. Operating Instructions
Table 3.2 (Continued) Math Menu
Menu Selection
Purpose/Action
Set Base
Sets selected trace as the base (denominator) trace
for math operations.
Difference
Creates subtraction trace from selected trace and
base trace (Difference = Selected – Base).
1st Derivative
Creates first derivative trace from selected trace.
Der.
Smooth Off/On
Toggles first derivative trace smoothing.
3.9.3. SELECT Button
The
SELECT
button switches which trace on-screen is the active trace. See Section 3.11.6 for
more information on selected traces.
3.9.4. AUTOFIT/HELP Button
The
AUTOFIT/HELP
button calls up the online help library and
AUTOFIT
soft-menu
options. Use the
M-FUNCTION
knob and
Select
menu option to scroll through and select the
help library content. When you select the
AUTOFIT
menu option, the CT100B looks for the
TDR signature of an open or short-terminated cable and selects appropriate horizontal and
vertical scale values to display the entire cable on the screen. By default the cursors are
positioned approximately to the beginning and end of the cable.
If
Vertical Recenter
is selected then only vertical scale and position are affected.
3.9.5. CURSOR Button
Toggles between cursors, making the active cursor inactive and the inactive cursor active.
Pressing and holding the
CURSOR
button for one second then releasing will reposition any
cursors not visible on the screen to the screen.
3.9.6. FILE Button and Menu
Use this menu to access the user library of configurations, cable types, and cable records.
FILE
button menu selections are shown in Table 3.3.
Table 3.3.
FILE Button and Menu
Menu Selection
Purpose/Action
Custom Cable Types
Use to add, select, or delete an entry from the
custom cable type library:
Add New ..........................
Enter in new cable type information.
Delete............................
Delete selected cable type from the library.
Select............................
Select the highlighted custom cable type. Instrument
settings change to reflect the cable Vp setting.
CT100B TDR Cable Analyzers Operator’s Manual
19