![Mohr CT100B Series Скачать руководство пользователя страница 121](http://html.mh-extra.com/html/mohr/ct100b-series/ct100b-series_operators-manual_1823448121.webp)
C. Operator Troubleshooting
C.3.10
File IO Verification
Attach a USB
Keyboard to the
front of the device
While in the SCAN
menu select SAVE.
In the Save Trace
dialog, use the
keyboard to enter a
name for the trace.
Does the
keyboard work?
NO
FAILED
YES
After entering a
name for the
scanned trace, press
OK
Did the trace save
successfully?
Navigate to File →
Cable Scan Records
NO
Does cable scan
records list a
collection of saved
traces?
Export the traces
using a USB drive or
CT Viewer and then
delete the traces on
the device
YES
Attempt to save the
scanned trace again.
Did the trace fail
to save due to
insufficient space?
YES
YES
Hide the saved trace and
attempt to reload the
saved trace. See section
3.11.8
for instructions on
loading a saved trace.
NO
NO
Did the trace
load correctly?
NO
YES
PASSED
Verify that only the live TDR trace is on
the display. Enclose the leading edge of
the live trace with the two cursors and
generate a cursor scan. See section
3.11.5
for instructions on creating a scan.
START
CT100B TDR Cable Analyzers Operator’s Manual
103