Convenience Handle
1. Attach the convenience handle track anchor to the vertical track adjacent to the brake release handle
(Figure 13a).
2. Attach the convenience handle to the anchor and press the screw covers into the handle (Figure 13b).
Figure 13a.
Figure 13b.
D
Detail D
C
Detail C
MEDICAL GAS CONNECTIONS
1. Use extreme care in making the medical gas service connections.
2. Each connection has been provided with a color-coded medical gas label.
3. All medical gas connections are to be made and tested in accordance with NFPA 99.
4. Do not connect the medical gas hoses from the column below to the DISS checks on the service entry bracket
until all brazed connections and testing of medical gas piping systems have been completed.
ELECTRICAL CONNECTIONS
1. Ensure that all electrical power circuits are locked off prior to hook-up.
2. Review the wiring diagram or shop drawings.
3. All connections are labeled on the unit at the point of connection.
4. Connect conduits and wiring from the column below to the electrical junction boxes on the service entry
bracket and make connections inside the boxes, as required. Make connections on the brake control harness
on both the column end of the unit and to the brake power supply enclosure.
NOTE: The power supply for the brake circuit has an auto-sensing input of 100-120VAC/200-240VAC, 50/60Hz.
5. Test electrical connections in accordance with state and local codes.
LOW-VOLTAGE CONNECTIONS
1. Install each low-voltage device in accordance with manufacturer’s installation instructions.
2. Test equipment in accordance with state or local codes.