971 UniMag
™
Operation manual
14
The setpoint safety delay function requires 5 continuously measurements that exceeds setpoint value before
the relay is tripped. This feature prevents that noise or pressure pulse can trig the relay. If fast setpoint
response is required the setpoint safety delay can be disabled.
Setpoint safety delay
Command:
@253SPD!ON;FF
Command values:
ON, OFF
Command reply:
@253ACKON;FF
Query:
@253SPD?;FF
Query reply:
@253ACKON;FF
Factory default:
ON
The 971 with integrated display provides real time pressure readout, pressure trend and setpoint status.
The display setup can be changed via the intuitive menu interface. The settings menu views transducer
settings.
Use the [
] and [
] keys to scroll the cursor and select menu point by pressing the [+] key.
Unit:
Select pressure unit Torr, mbar or Pascal. (Will not change transducer unit setup)
Orientation: Select display orientation
Settings:
View transducer parameters
Trend display
Pressure reading
Exponent
Sensor reading
Setpoint Status
Pressure unit
Green=Activated, Red=Deactivated
Main menu
Unit selection
View Settings
Содержание UniMag 971
Страница 1: ...P N 100017131 971 UniMag Transducer Operation and Installation Manual Revision C January 2010...
Страница 36: ...971 UniMag Operation manual 33 mm Inch mm Inch...
Страница 37: ...971 UniMag Operation manual 34 mm Inch...
Страница 40: ...971 UniMag Operation manual 37...
Страница 41: ...971 UniMag Operation manual 38...
Страница 42: ...971 UniMag Operation manual 39...
Страница 44: ...P N 100017133 971 UniMag Transducer Operation and Installation Manual Revision C January 2010...