 File: SM1280_r7eng - April 2017 - Technical note: SM1280

             

SM 1280

 

This is the set-mode submenu: it 
allows to select Manual or Exter-
nal Control.
In Manual mode the power can be 
adjusted using the panel knob. In 
External Control mode the manual 
commands are not effective.

This is the set-efficiency 

submenu: it allows to enter the 
estimated value of magnetron 

efficiency, range: 0-100% . If 

you select 100% the “Microwave 
Power” corresponds to the electri-
cal output power of the psu.

When the unit is in Alarm state 
the cause is displayed until the 
alarm is removed.

The CPU drives the unit to  obtain an electrical output power 
(Va * Ia) value proportional in a linear way to the reference 
input signal. The scaling is 850W/V.

INTERLOCK FUNCTION:  managed throu CONN# 4, pin 5 - 
6, “MICROSWITCH ON COVER”. It is an external hardware 
chain of contacts, that must be closed all at the time to have 
the internal relay closed. (ON condition). When in ON the 
CPU light-up the related Interlock OK led; in OFF state the led 
is Off and the unit goes in Stand-by, as all dangerous voltages 
on the head connector are off.

HEAD ENABLE 1, 2: CONN# 4, pin 7 - 9; CONN# 6, PIN 
10 - 23 not implemented yet (version 1.0.1). Do not use, they 
have no meaning.

Firmware upgrade

The software stored on the CPU (firmware, on a flash 

memory) can be upgraded to a newer version using the RS232 
serial port on the unit, connected to a PC.

Important remark: at power-on, even if a program is already 
stored, the microprocessor enters a “BOOTLOADER” status, 
lasting 2 seconds, during which it is possible to overwrite, 
through the serial RS232 port, the stored code.
If no new code is loaded within 2 seconds, the previously stored 
one is run.

To upload a new program through the bootloader, simply 
connect the SM1280 serial port to the PC serial port, using a 9 
pin 1:1 cable (only TX, RX and Ground are used, but also a full 

cable will fit. See schematic on page 14)  

No external switch must be operated, and the unit does not start 
until the new program is stored, but be sure no enable signal is 
active when upgrading.

The upgrade of the firmware embedded in the SM1280 is 

released as a .zip archive named “

program_name.zip

”.

Unpacking the .zip archive you generate:
- a directory named “

program_name

” containing the micro 

code to be loaded, 
- an executable program named “

SendApplicationProgram.bat

”- and the loader “

sRecLoad.exe

”.

Once unpacked the .zip program
- enter in the folder “

program_name

”, 

- switch-on the SM1280
- run 

whithin 2 seconds

 the program “

SendApplicationProgra-

m.bat

” (which launches 

sRecLoad.exe 

application)

(This is the maximum time that the microprocessor bootloader 
waits for new code to be loaded).

During the bootloader wait time, the red ALARM led is on: if 
no code is sent, then the previous code is executed, and it is not 
possible to change it without switching the unit off.

Meaning of external commands.

(vers 0 - 2 - 20 - 50)

HIBERNATE: has value ON/OFF. If ON the magnetron’s 

filament is not powered and the SM1280 goes into Stand-by 

mode.

In OFF the unit makes filament heating and, when the related 

timer is elapsed, enters in Ready mode. 
If a user don’t like to use the Hibernate function, he may not 
connect the HIBERNATE control signal: the system then 
detect it as in OFF condition.

POWER ENABLE: has value ON/OFF, but its effect is related 
also with the Hibernate command as follow:
a) in OFF state - and HIBERNATE is OFF - the unit performs 

filament preheating and goes in Ready mode but does not 

generate output power, even if the reference signal is >0;
b) in OFF state - and HIBERNATE is ON - the unit is in 
Stand-by mode but does not generate output power, even if the 
reference signal is >0;
c) in ON state - and HIBERNATE is OFF - the unit waits the 
end of preheating timer, if not yet elapsed, then starts  to gene-
rate output power;
d) in ON state - and HIBERNATE is ON - the unit is in Stand-
by mode but does  not generate output power, even if the 
reference signal is >0.

POWER SET: it’s an analog input value and has the meaning 
of  “desired percent” of the full output power (8500W)
When this value is 

lower than 1Vdc the unit does not gene-

rate any output power,

 when >1Vdc the unit generates power 

proportional to this signal value.

Содержание SM 1280

Страница 1: ...SM1280D for line 3 x 208V SM1280T for line 3 x 400V SM1280U for line 3 x 480V SWITCHING POWER GENERATOR WITH EMBEDDED MICROCONTROLLER FOR 6 KW MAGNETRON TECHNICAL NOTE issue April 2017 SM 1280...

Страница 2: ...Cover Rear...

Страница 3: ...ioni degli Stati membri relative al materiale elettrico destinato ad essere adoperato entro taluni limiti di tensione 2014 30 UE Harmonisation of the laws of the member States relating to electromagne...

Страница 4: ...ollowing Directive rules this WEEE is not a municipal waste and should be collected separately according to rules of your country the symbol means that this WEEE cannot be disposal in the same place a...

Страница 5: ...d by persons acquainted with the regulations covering the application The equipment described in this manual has been classified as industrial scientific and medical ISM radio frequency equip ment and...

Страница 6: ...t provide a proper cooling flow the cooling air enter into the unit from the front side it passes through the internal circuits and components then it goes out from the back The intake air flow is app...

Страница 7: ...ICH EINEN SCHLOG HOLEN ATTENZIONE LA MANUTENZIONE DEVE ESSERE FATTA DA PERSONALE QUALIFICATO RISCHIO DI SCARICHE ELETTRICHE ALL INTERNO COLLEGARE ALLA TERRA DEL GENERATORE REMOTO VERBINDER DIRECT MIT...

Страница 8: ...V output 3 because the SM1180 had a 4 pole LH type connector and the SM1280 uses the single pole GES type HV connector The back panel of vers 22 see below has the same connetor style size and pin out...

Страница 9: ...Frequency kW MHz 6 2450 Typical YJ1600 H3881 PM60 AM3881 E Selectable filament backdown curve for the specific tube Enabling signal digital input Vdc 24 Range 5 30 ON OFF output alarm reset Input Z 4k...

Страница 10: ...for industrial use only not for laboratory use and user must respect the wiring norms and prescriptions as described into next paragraph Wiring instruction The equipment cannot operate cantilevered i...

Страница 11: ...nnector is not intended for live insertion of the unit an external tripolar breaker must always be pro vided and the unit must be switched off before inserting unplugging of the connector IV The conne...

Страница 12: ...e unit enters in Stand By mode stand by it s a waiting state the unit is powered but not ready as the magnetron filament is not powered all the wor king parameters can be setted from Operator Panel th...

Страница 13: ...e Display versions have a graphic display can be driven manually via the front panel controls The status of the equipment and all the relevant information to the user are provided through the graphic...

Страница 14: ...new program is stored but be sure no enable signal is active when upgrading The upgrade of the firmware embedded in the SM1280 is released as a zip archive named program_name zip Unpacking the zip ar...

Страница 15: ...g the head in a safe condition SW1 8 not used SW2 NORMAL OP HARDWARE FIXED TEST FIELDBUS PM60 H3881 ON OFF ON OFF ON OFF ON OFF YJ1600 ON OFF id26042234 pdfMachine by Broadgun Software a great PDF wri...

Страница 16: ...units A value from 1 to 99 can be set while node 00 is usually reserved and must not be used consult the specific software documentation The bus is based on a network interface device named AnyBus Com...

Страница 17: ...ng the output in the state they were On contrary if the Process Active TimeOut value consult the software documentation for the specific bus has been set to a value greater than zero it stays in Proce...

Страница 18: ...page 16 of 27 File SM1280_r7eng April 2017 Technical note SM1280 SM 1280 Connector 1 wiring vers 0 1 2 22...

Страница 19: ...File SM1280_r7eng April 2017 Technical note SM1280 page 17 of 27 SM 1280 SM 1280 Connector 1 wiring vers 50 51...

Страница 20: ...page 18 of 27 File SM1280_r7eng April 2017 Technical note SM1280 SM 1280 Connector 3 4 5 wiring all versions except 22 14 13 12 11 7 3 8 4 1 2...

Страница 21: ...rated differently i e the reverse signal may be scaled 0 6000W Forward and Reverse Power Monitor if signals are availables Magnetron voltage and current Monitor default setting CONN 6 Forward Power 10...

Страница 22: ...17 Technical note SM1280 SM 1280 RS232 connector wiring all versions id32273156 pdfMachine by Broadgun Software a great PDF writer a great PDF creator http www pdfmachine com http www broadgun com XX...

Страница 23: ...File SM1280_r7eng April 2017 Technical note SM1280 page 21 of 27 SM 1280 SM 1280 Connector 3 4 5 wiring for version 22 only Typical wiring to our standard head type TM060 61...

Страница 24: ...page 22 of 27 File SM1280_r7eng April 2017 Technical note SM1280 SM 1280 COMPONENT LAYOUT Vers 0 1 2 50 51...

Страница 25: ...10 8 1 Aux contact IA3RV1901 1B 9 1 Main line filter FL50FCD10 10 1 Auxsiliary filter FLFN20601206 11 1 3 pins Socket AMP CPC MOCPC07 2060362 CONN 7 3 male pin AMP type XII 35A MOCPCX 0662614 CONN 7 1...

Страница 26: ...s Plug AMP PLUG 1 7 female pin type XII 35A PLUG 1 pins female pin type XII 60A 22 1 HV inductor 23 1 LV inductor 24 1 Current feedback board M302 800000302 1280 25 1 Air grid 26 1 Field bus board M36...

Страница 27: ...ted if true check for h v spot discharge inside socket plug in case replace them d The filament is cold check filament current with high voltage OFF enable signal OFF h v connector 3 disconnected Chec...

Страница 28: ...ad in case replace the magnetron 10 Rack Error The unit has detected an output current 10 lower than expected usually due by a sudden drop of the main line voltage check the main line 11 Not implement...

Страница 29: ...the input for auxiliary voltage separated from power input with bus like version 1 20 special direct replacement for customized SM1180 300 produced till 2001 22 legacy direct replacement for SM1180 a...