Mini-Convectron Module Instruction Manual - 275513
53
Appendix A Specifications
Pressure Measurement
Measurement Range for Air or N
2
Torr
1 x 10
–4
to 1000
mbar
1 x 10
–4
to 1333
Pascal
1 x 10
–2
to 1.33 x 10
5
Resolution
Torr
1 x 10
–4
mbar
1 x 10
–4
Pascal
1 x 10
–2
Measurements will change with different gases and mixtures. Do not use the module with flammable or
explosive gases.The module is factory calibrated for use on N
2
. It also measures the pressure of air correctly
within the accuracy specification for the instrument. If the module will measure the pressure of a gas other
than N
2
or air, you must adjust relays for the process gas.
Temperature Limits
Operating Temperature
+0 to +40 °C (+32 to +104 °F) ambient, non-condensing
Non-operating Temperature
–40 to +70 °C (–40 to +158 °F)
CE Compliance
EMC
EN61326-1
Safety
EN61010-1
IP Rating
IP20
Power Requirements and Electrical Connections
Power Requirements
With optional display
11.5 to 26.5 Vdc, 0.12 A at 11.5 Vdc, 1.8 W maximum
Without display
11.5 to 26.5 Vdc, 0.1 A at 11.5 Vdc, 1.2 W maximum
I/O connectors
2 setpoint relays
15-pin male, high-density subminiature D
1 setpoint relay
9-pin male, subminiature D
Содержание Series 275 MiniConvectron
Страница 2: ......
Страница 6: ...Table of Contents 6 Mini Convectron Module Instruction Manual 275513...
Страница 52: ...Chapter 4 52 Mini Convectron Module Instruction Manual 275513...
Страница 56: ...56 Mini Convectron Module Instruction Manual 275513...
Страница 58: ...58 Mini Convectron Module Instruction Manual 275513...
Страница 62: ...Index 62 Mini Convectron Module Instruction Manual 275513...
Страница 63: ......