Chapter Four: Operation
How To Adjust the Zero
43
Chapter Four: Operation
How To Adjust the Zero
All pressure transducers require initial and periodic zero adjustments. Prior to initial operation and
during periodic maintenance you must check the transducer zero to verify the proper output. The
zero can be set (or reset) by adjusting the
ZERO
potentiometer on the side of the transducer
(refer to Figure 1, page 25, and Figure 3, page 27) or at the front panel of any MKS Power
Supply/Readout being used.
To achieve the full dynamic range specified for the transducer, the zero adjustment
must
be made
at a pressure one decade lower than the transducer’s minimum resolution (0.001% of FS).
Zeroing a transducer above its stated minimum resolution creates a
zero offset
relative to true
absolute pressure. All subsequent readings are then linear and accurate
relative to the offset
value
.
Note
If available pressures are not sufficiently low to set the transducer zero,
you may use a vacuum leak detector with sufficient vacuum pumping (to
achieve proper zeroing pressures). In this case, mount the transducer on
the leak detector
in the same plane of orientation as it will be during
actual use
.
Содержание Micro-Baratron 870B
Страница 3: ...123655 P1 Rev B 06 01 MKS Type 870B 872B Micro Baratron Absolute Pressure Transducer...
Страница 8: ......
Страница 10: ......
Страница 28: ......
Страница 52: ...Labels Chapter Three Overview 42 This page intentionally left blank...
Страница 58: ...Troubleshooting Chapter Five Maintenance and Troubleshooting 48 This page intentionally left blank...
Страница 62: ...Environmental Specifications Appendix A Product Specifications 52 This page intentionally left blank...
Страница 66: ...Model Code Appendix B Model Code Explanation 56 This page intentionally left blank...