Connectors
Chapter Two: Installation
34
Flying Leads
The 870/872 transducer is available with a flying lead at the end of either a 6’ or 10’ cable (refer
to Figure 9). The two cable lengths have different configurations, as noted in Tables 8 and 9.
6' (1.8 m)
Model Code "F"
10' (3 m)
Model Code "L"
or
Black
Red
White
Green
Transducer Body
Braided Shielded Cable
Figure 9: Flying Leads
Table 8 lists the pinout for the Flying Lead with a 6’ (1.8 m) cable.
Flying Lead Pinout (Model Code F)
Color
Voltage Signal
Current Signal
Red
Pressure Output
Do Not Use
Black
Pressure Return
Do Not Use
White
Power Return
4 to 20 mA negative excitation
Green
+ Power Input
4 to 20 mA positive excitation
Table 8: Flying Lead Pinout (Model Code F)
Table 9 lists the pinout for the Flying Lead with a 10’(3 m) cable.
Flying Lead Pinout (Model Code L)
Color
Voltage Signal
Current Signal
Red
+ Power Input
4 to 20 mA positive excitation
Black
Power Return
4 to 20 mA negative excitation
White
Pressure Return
Do Not Use
Green
Pressure Output
Do Not Use
Table 9: Flying Lead Pinout (Model Code L)
Содержание Micro-Baratron 870B
Страница 3: ...123655 P1 Rev B 06 01 MKS Type 870B 872B Micro Baratron Absolute Pressure Transducer...
Страница 8: ......
Страница 10: ......
Страница 28: ......
Страница 52: ...Labels Chapter Three Overview 42 This page intentionally left blank...
Страница 58: ...Troubleshooting Chapter Five Maintenance and Troubleshooting 48 This page intentionally left blank...
Страница 62: ...Environmental Specifications Appendix A Product Specifications 52 This page intentionally left blank...
Страница 66: ...Model Code Appendix B Model Code Explanation 56 This page intentionally left blank...