14
Digital Pirani vacuum Sensor System
Relay Inductive Loads and Arc Suppression
If the set point relay is used to switch inductive loads, e.g., solenoids, relays,
transformers, etc., the arcing of the relay contacts may interfere with
Controller operation or reduce relay contact life. Therefore an arc suppression
network, shown schematically below, is recommended. The values of the
capacitance
C and the resistance R are calculated by the equations,
C = I
2
/(1 x 10
7
) and
R = E/10I
a
where,
C is in farads
R is in ohms
I is DC or AC
peak
load current in amperes
E is DC or AC
peak
source voltage in volts
a = 1 + (50/E).
Note that,
R
min
= 0.5
Ω
and
C
min
= 1.0 x 10
-9
F.
Pin
Description
1
Set point relay 1 - normally closed contact
2
Set point relay 1 - normally open contact
3
Set point relay 2 - common
4
Not used
5
Analog output voltage (+)
6
Set point relay 1 - common
7
Set point relay 2 - normally closed contact
8
Set point relay 2 - normally open contact
9
Analog output voltage (-)
9-pin Accessory port
1
5
9
6
E
R
C
Содержание HPS 945 Series
Страница 1: ...HPS Products Series 945 Digital Pirani Vacuum Sensor System OPERATION AND MAINTENANCE MANUAL...
Страница 2: ......
Страница 32: ...26 Digital Pirani vacuum Sensor System Notes...
Страница 35: ...A 3 Digital Pairani Vacuum Sensor System Notes...
Страница 36: ...A 4 Digital Pairani Vacuum Sensor System...
Страница 37: ......
Страница 38: ......