10

Pirani Vacuum Sensor System

To fit the KF 16 port to a KF 10 port, use an HPS

TM

 adaptive centering ring

(see Accessories, p. 21).

To fit the Sensor with the 

1

/

8

" NPT, 

do not use the case for tightening. The

Sensor’s vacuum tubing has 

9

/

16

" hex flats for tightening. Wrap some Teflon

®

tape on the threads of the Sensor, in the direction of installation, to ensure a
leak-free seal.

Instead of the NPT fitting, you may opt to use the ½" O-ring compression
seal against the vacuum tubing above the thread, 

but the O-ring seal cannot

be used for positive pressure applications.

A solid electrical connection between the Sensor and the
grounded vacuum system must be provided to shield the
tube element from external radiation sources. In applications
where the system may be exposed to large voltage
fluctuations, a centering ring with a screen should be
installed, and the screen

and tubing then grounded.

Pirani Controller Installation

Controller Mounting

The Series 917 Pirani Controller is designed for either panel mounting or for
stand-alone use. Regardless of the method you choose, assure adequate
ventilation to the Controller with at least a ½ inch left open above and below it.

Adhesive backed rubber feet are provided for bench top use. Remove the
adhesive backing from each foot and apply them to each corner of the
aluminum bottom surface.

An optional hardware kit is also available for mounting the Controller into a
standard ¼ DIN cutout in a panel up to 

3

/16-inch thick (see Accessories,

p. 21). A dimensioned illustration shows the required cutout on the top of the
next page. Leave at least 3 inches of clearance behind the Controller's rear
panel to accommodate the connectors and cables.

Mount the Controller into the panel by slipping it through the cutout from the
front. Slide the panel mounting brackets into the slots on either side of the
Controller from the rear, and secure them with the thumbscrews provided.

Electrical Connection of the Pirani Sensor to the Controller

Connect the Pirani Sensor to the Controller with a sensor cable  (see
Accessories, p. 21). This cable has a standard octal socket. With an
integrated polarizing tab, the socket has sufficient contact pressure so that
strain relief is unnecessary in most applications.

Содержание HPS 917

Страница 1: ...HPSTM Products Series 917 Pirani Vacuum Sensor System OPERATION AND MAINTENANCE MANUAL...

Страница 2: ......

Страница 3: ...Pirani Vacuum Sensor System HPS Products Series 917 Pirani Vacuum Sensor System AUGUST 1999 PART 109170020 REV B...

Страница 4: ...l number is located on the back panel of the Pirani controller 1994 1996 by MKS Instruments Inc HPSTM Products Inc All rights reserved SensaVac is a trademark of MKS Instruments Inc Teflon is a regist...

Страница 5: ...allation 10 Controller Mounting 10 Electrical Connection of the Pirani Sensor to the Controller 10 AC Power Cord 11 Fuse Replacement 12 Accessory Connector 12 Relay Inductive Loads and Arc Suppression...

Страница 6: ...Pirani Vacuum Sensor System...

Страница 7: ...er immediately Keep all shipping materials and packaging for claim verification Package Contents The first two symbols below located on your Series 917 Controller and or Sensor identify critical safet...

Страница 8: ...r gases Take precautions to avoid overpressure during systembackfill When backfilling do not use the Series 917 Pirani System to read pressures above 100 Torr In a system which is filled with a pressu...

Страница 9: ...and current rating Only qualified personnel should change line voltage or fuse ratings if required Outside of the U S use only a harmonized detachable cord set that has conductors with a cross section...

Страница 10: ...eRange 10o to 55o C 14o to 131o F OperatingHumidity 95 maximum noncondensing InstallationCategory Overvoltage Category II PollutionDegree2 Number of Channels 1 Front Panel Controls 2 nonvolatile indep...

Страница 11: ...aterials Exposed to Vacuum 304 stainless steel platinum alumina ceramic silver brazing alloy nickel 200 Internal Volume 0 5 in 3 8 0 cm3 maximum OperatingTemperatureRange 0o to 50o C 32o to 122o F Max...

Страница 12: ...Point Indicators Power Cord Inlet w Fuse Holder Panel Mounting Fastener Holes Male 15 pin Accessory Port Female 15 pinGauge Port Slots for Optional Mounting Brackets Male Octal Socket Sensor Vacuum Po...

Страница 13: ...own using the relay set points Sensing abnormal pressure and taking appropriate security measures using the relay set points Controlling system pressure using the analog output as input to an automati...

Страница 14: ...iently accessed from the front panel and cable connections are hidden to the rear The Series 917 Pirani Controller is compatible with any HPSTM standard Pirani Sensor without loss of calibration accur...

Страница 15: ...ght affect the tube s element For example if the Sensor is installed directly above a roughing pump in the system oil vapor could contaminate the tube s filament wire and cause the emissivity and the...

Страница 16: ...or stand alone use Regardless of the method you choose assure adequate ventilation to the Controller with at least a inch left open above and below it Adhesive backed rubber feet are provided for benc...

Страница 17: ...engths up to 500 feet Connect the cable to the rear of the Controller at the D sub port labeled Gauge Tighten the connector jackscrews into the mating screwlocks to ensure proper electrical connection...

Страница 18: ...1 A 5 mm x 20 mm 90 to 130 VAC T 0 04 A 5 mm x 20 mm 210 to 250 VAC 4 Close and secure the fuse holder AccessoryConnector Relay set point contacts and the analog output voltage can also be accessed f...

Страница 19: ...uppression If a set point relay is used to switch inductive loads e g solenoids relays transformers etc the arcing of the relay contacts may interfere with Controller operation or reduce relay contact...

Страница 20: ...n the unit Turn the power on to display pressure or to operate the set point relays The graph below shows the Series 917 Pirani System s voltage output as a function of pressure for nitrogen helium an...

Страница 21: ...330 2 0E 02 2 7E 00 0 422 0 340 0 371 3 0E 02 4 0E 00 0 491 0 400 0 429 4 0E 02 5 3E 00 0 552 0 444 0 481 5 0E 02 6 7E 00 0 606 0 485 0 527 6 0E 02 8 0E 00 0 657 0 520 0 569 7 0E 02 9 3E 00 0 702 0 55...

Страница 22: ...inherent sensitivity to gas type makes it useful for detecting leaks It is a useful complement to a mass spectrometer leak detector which is most efficient in locating smaller leaks The Series 917 Pi...

Страница 23: ...not recommended Leak testing can be done with a mechanical pump however they may cause cyclical variations in pressure with rotation of the vanes This shows up as a large background noise signal possi...

Страница 24: ...n indicated This reading error could lead to a dangerous overpressure A system which is backfilled from a pressurized gas source should contain a safety device such as a burst disc Calibrating for Oth...

Страница 25: ...e The Series 917 Pirani Controller is designed to be maintenance free under normal operation In the unlikely event that problems occur this section lists symptoms and associated possible causes With t...

Страница 26: ...with an ohmmeter with less than 5 mA of current The resistance readings of a normal Sensor measured at atmospheric pressure and at room temperature 20 C are shown here Cleaning the Pirani Sensor Rough...

Страница 27: ...103150012 1 8 NPT M with Compression 103150011 Seal Option 15 mm x 30 mm Tubing 103150016 18 mm x 30 mm Tubing 103150018 Power Cord 115 VAC 103150001 Seal Centering Ring Assemblies with Screen KF 16...

Страница 28: ...id or unpaid service is desired if the product is contaminated with potentially corrosive reactive harmful or radioactive materials gases or chemicals When products are used with toxic chemicals or in...

Страница 29: ...23 Pirani Vacuum Sensor System Notes...

Страница 30: ...24 Pirani Vacuum Sensor System...

Страница 31: ......

Страница 32: ......

Отзывы: