Series 203 Variable Leak Valve Installation and Operating Instructions

1

Series 203 Variable Leak Installation
and Operating Instructions

1

Safety  Hazards

2

Important Precautions

The Variable Leak is a carefully designed and manufactured variable 
conductance device. With proper care and use it will give long and trouble 
free service under the most extreme conditions. The following are important 
precautions which must be observed at all times to prevent possible 
irreparable damage or poor performance.

2.1 DO NOT remove the valve from its container until these instructions 

have been read completely.

2.2 Always cover the ports with plastic cap plugs or rubber stoppers and 

place the entire unit in a clean polyethylene bag whenever the Leak is 
stored, shipped or not connected to the vacuum system. Do not use 
cork stoppers or adhesive tape.

2.3 Carefully exclude dust, dirt, chips, moisture and other contaminants. A 

liquid seal, formed from condensed vapors on the sealing surface, will 
result in loss of leak control. Foreign or corrosive material will 
permanently damage the sealing surfaces.

WARNING

Use of certain corrosive gases can cause gases to leak 
through the sealed valve or into the atmosphere, 
resulting in serious personal injury.

Before using a gas with the valve, check the gas for 
compatibility with the following materials:

• Fine  silver

• Type 304S ST

WARNING

Exceeding the maximum internal pressure specified 
below can cause an explosion, resulting in property 
damage or serious personal injury.

• Inlet pressure should never exceed 200 psig.

• For inlet pressures less than one atmosphere, connect the gas 

source to either port.

• For gas pressures greater than one atmosphere, connect the gas 

source to the center port.

Содержание Granville-Phillips 203 Series

Страница 1: ...Series 203 Instruction Manual Granville Phillips Variable Leak Valve Instruction manual part number 203026 Revision B November 2016...

Страница 2: ......

Страница 3: ...red trademarks are the properties of their respective owners Customer Service Technical Support MKS Pressure and Vacuum Measurement Solutions MKS Instruments Inc Granville Phillips Division 6450 Dry C...

Страница 4: ......

Страница 5: ...1 Conductance 3 3 2 Throughput 3 3 3 Maximum Inlet Pressures 3 3 4 Gases Controlled 3 3 5 Materials 3 3 6 Temperature 4 3 7 Vacuum Connections 4 4 Installation 4 5 Operation 5 6 Removing the Driver 5...

Страница 6: ...ii Series 203 Variable Leak Valve Installation and Operating Instructions...

Страница 7: ...ped or not connected to the vacuum system Do not use cork stoppers or adhesive tape 2 3 Carefully exclude dust dirt chips moisture and other contaminants A liquid seal formed from condensed vapors on...

Страница 8: ...5 High temperature bake the Leak valve body only when the Driver Assembly is removed and the Leak held open with a bake out clamp 2 6 Stable ambient temperature is necessary to maintain conductance at...

Страница 9: ...flow rates some gas will continue to emerge from the seal for several minutes even though the seal is mass spectrometer leak tight This is a normal condition with metal to metal seals 3 2 Throughput...

Страница 10: ...lable are versions using Cajon VCR and VCA fittings 4 Installation Use every precaution to prevent contamination from entering the Leak Keep plastic caps or rubber stoppers on the valve ports until th...

Страница 11: ...t be removed for bakeout 6 1 Remove the name plate 6 2 Turn the driver handle until the counter reads approximately 150 digits above the previous sealing point to relieve all thrust on the nosepiece 6...

Страница 12: ...n 30 and 50 inch ounces torque If a leak detector is not available close the valve in small increments until a satisfactorily low conductance is attained 8 6 With the driver in the sealed position rem...

Страница 13: ...ings Relubricate the contact surfaces of the worm and gear and drive screw with molybdenum disulfide Reassemble Remove the retracting screw Loosen the set screw in the aluminum bushing and rotate one...

Страница 14: ...ided the item is returned to Granville Phillips Company together with a written statement of the problem 13 Specifications Materials Seal Pure silver All other interior surfaces 304 Stainless Steel Ph...

Страница 15: ......

Страница 16: ...l Support MKS Pressure and Vacuum Measurement Solutions MKS Instruments Inc Granville Phillips Division 6450 Dry Creek Parkway Longmont Colorado 80503 USA Tel 303 652 4400 Fax 303 652 2844 Email mks m...

Отзывы: