background image

Certification

8

Series 203 Variable Leak Valve Installation and Operating Instructions

11

Certification

MKS/Granville-Phillips certifies that this product met its published 
specifications at the time of shipment from the factory.

12

Limited Warranty

This Granville-Phillips product is warranted against defects in materials and 
workmanship for one year from the date of shipment provided the 
installation and preventive maintenance procedures specified in this 
instruction manual have been followed. Granville-Phillips Company will, at 
its option, repair or replace or refund the selling price of an item which 
proves to be defective during the warranty period provided the item is 
returned to Granville-Phillips Company together with a written statement of 
the problem.

13

Specifications

Materials

Seal

Pure silver

All other interior surfaces

304 Stainless Steel

Physical Data

Bakeout temperature

450 ° C, bakeout clamp attached
70 ° C valve open or closed driver installed

Seal deformation vs. number of closures

Negligible

Increase in closing torque vs. number of 
closures

Negligible

Flow path, flange to flange, when fully open

168 mm (6.62 in)

Maximum conductance for dry nitrogen at 
25 ° C

400 cc/sec.

Leakage conductance when fully closed

10

–11

 liters/sec.

Normal closing torque

50 inch-ounces

Standard connections

Type 304 Stainless Steel, Rotatable CuSeal or 
Mini-CuSeal flanges

Maximum Inlet Pressure

Connection type

Port A

Port B

CuSeal or Mini-CuSeal flange

25 psig

200 psig

Type 304 SS tubulation (welded to system)

25 psig

200 psig

Содержание Granville-Phillips 203 Series

Страница 1: ...Series 203 Instruction Manual Granville Phillips Variable Leak Valve Instruction manual part number 203026 Revision B November 2016...

Страница 2: ......

Страница 3: ...red trademarks are the properties of their respective owners Customer Service Technical Support MKS Pressure and Vacuum Measurement Solutions MKS Instruments Inc Granville Phillips Division 6450 Dry C...

Страница 4: ......

Страница 5: ...1 Conductance 3 3 2 Throughput 3 3 3 Maximum Inlet Pressures 3 3 4 Gases Controlled 3 3 5 Materials 3 3 6 Temperature 4 3 7 Vacuum Connections 4 4 Installation 4 5 Operation 5 6 Removing the Driver 5...

Страница 6: ...ii Series 203 Variable Leak Valve Installation and Operating Instructions...

Страница 7: ...ped or not connected to the vacuum system Do not use cork stoppers or adhesive tape 2 3 Carefully exclude dust dirt chips moisture and other contaminants A liquid seal formed from condensed vapors on...

Страница 8: ...5 High temperature bake the Leak valve body only when the Driver Assembly is removed and the Leak held open with a bake out clamp 2 6 Stable ambient temperature is necessary to maintain conductance at...

Страница 9: ...flow rates some gas will continue to emerge from the seal for several minutes even though the seal is mass spectrometer leak tight This is a normal condition with metal to metal seals 3 2 Throughput...

Страница 10: ...lable are versions using Cajon VCR and VCA fittings 4 Installation Use every precaution to prevent contamination from entering the Leak Keep plastic caps or rubber stoppers on the valve ports until th...

Страница 11: ...t be removed for bakeout 6 1 Remove the name plate 6 2 Turn the driver handle until the counter reads approximately 150 digits above the previous sealing point to relieve all thrust on the nosepiece 6...

Страница 12: ...n 30 and 50 inch ounces torque If a leak detector is not available close the valve in small increments until a satisfactorily low conductance is attained 8 6 With the driver in the sealed position rem...

Страница 13: ...ings Relubricate the contact surfaces of the worm and gear and drive screw with molybdenum disulfide Reassemble Remove the retracting screw Loosen the set screw in the aluminum bushing and rotate one...

Страница 14: ...ided the item is returned to Granville Phillips Company together with a written statement of the problem 13 Specifications Materials Seal Pure silver All other interior surfaces 304 Stainless Steel Ph...

Страница 15: ......

Страница 16: ...l Support MKS Pressure and Vacuum Measurement Solutions MKS Instruments Inc Granville Phillips Division 6450 Dry Creek Parkway Longmont Colorado 80503 USA Tel 303 652 4400 Fax 303 652 2844 Email mks m...

Отзывы: