Chapter Three: Overview
UGeneral Information
Chapter Three: Overview
General Information
A complete pressure transducer system requires three components to convert pressure to a linear
DC voltage output: a sensor, signal conditioner, and power supply. An analog or digital meter is
required to display the DC output in pressure units.
The 722B transducer contains two of the above components: the sensor and signal conditioner.
An MKS or MKS-compatible power supply is required to complete the pressure to DC voltage
output conversion, and an MKS or MKS-compatible display unit is required for direct pressure
readout. The display unit could be a personal computer, an MKS pressure controller, or an MKS
PDR Series power supply/readout unit.
Sensor
The variable capacitance sensor consists of a pressure inlet tube (port) connected to a small
chamber in the transducer body. One wall of this chamber is a metal diaphragm. The front side
of the diaphragm is exposed to the gas whose pressure is to be measured. The back, or
reference
side of the diaphragm faces a rigidly mounted ceramic disc containing two electrodes. The
diaphragm is positioned opposite the inlet port.
The reference side is permanently evacuated below the resolution of the instrument and its
vacuum is maintained with a chemical getter system.
The diaphragm deflects with changing pressure (force per unit area) independently of the gas type
or composition of the measured gas. This deflection causes an imbalance of the sensor electrode
capacitances since the distance to the diaphragm is now different for each electrode. Using a
precision constant frequency oscillator for excitation, the imbalance of capacitances is converted
to a DC voltage. The resultant signal is then linearized, zeroed, and amplified via the signal
conditioner electronics, to produce a precise output signal scaled to the range of the transducer.
59
Содержание Baratron 722B
Страница 1: ...1034138 001 Revision A 5 09 MKS Baratron Type 722B Absolute Pressure Transducer...
Страница 5: ...Table of Contents NW 16 KF Fitting Information 69 Index 71 v...
Страница 6: ......
Страница 8: ...List of Figures and Tables viii...
Страница 58: ...This page intentionally left blank 58...
Страница 62: ...This page intentionally left blank 62...
Страница 70: ...This page intentionally left blank 70...