52
Electrical Information
The 722B transducer requires an external power source capable of supplying the voltages listed in
Appendix A: Product Specifications,
page 65. Noise and ripple should be less than 20 mV
(peak-to-peak) over a 10 kHz bandwidth. You may use any readout device capable of reading
from -0.6 V to 11 V. Refer to Figure 1 for the power, signal and chassis grounding scheme for a
voltage unit.
Note
The ground of any external power supply and readout should be the same
as the transducer ground (chassis ground) to minimize any possible
ground loops and power supply noise which can affect the performance
and stability of the system.
Readout
Power
Supply
+
-
0 to 10 V Pressure Output
Pressure Return
Power Return
10K
Note 2:
The absolute value of the potential difference between the chassis ground and
signal common/power ground should not exceed 14 Volts.
Note 1:
For best results, use a readout with fully differential inputs.
Multiple Ground
Tie-Point
0.3 F
µ
+13 to +32 V (for 0 to 10 V)
Chassis
Ground
Sensor Port
Figure 1: Electrical Scheme for a Voltage Unit
NW16-KF Fitting Information
The NW16-KF fittings are only available for units with full scale pressure of a maximum of 5000
Torr (100 psia) or less.
Warning
Units with NW16-KF fittings and a full scale range greater than
1200 Torr (23 psi) require an HPS overpressure ring.
Operating the unit without a protective overpressure ring may
result in injury.
The HPS part number for the overpressure ring is 100316301.
Содержание Baratron 722B
Страница 1: ...1034138 001 Revision A 5 09 MKS Baratron Type 722B Absolute Pressure Transducer...
Страница 5: ...Table of Contents NW 16 KF Fitting Information 69 Index 71 v...
Страница 6: ......
Страница 8: ...List of Figures and Tables viii...
Страница 58: ...This page intentionally left blank 58...
Страница 62: ...This page intentionally left blank 62...
Страница 70: ...This page intentionally left blank 70...