Chapter Two: Installation
U How To Check the Transducer Zero
55
Attaching the Terminal Block Connector Cable
The cable to the terminal block connector must be firmly attached to the top of the transducer, in
order to comply with CE Mark requirements. Figure 2 shows the preferred method to connect the
cable; using a metal cable clamp to affix the cable to the transducer. Figure 3 shows an alternate
method; wrapping the braided shield under the screw. Both examples show the cable securely
clamped to the transducer.
Transducer
Overall Insulation
(if present)
Bare Metal Cable Clamp
Making Firm Contact To Braid
Braid Here Is Desirable
(but not usually necessary)
Keep Wires and Braid
Flat Against Case
Metal Cable Clamp
Screw
Split Lock Washer
External Tooth Lock Washer
Transducer Housing
Optional Plastic or Metal Cable
Clamp (For Physical Strain Relief)
Figure 2: How To Connect the Cable to a Terminal Block (Example 1)
Transducer
Overall Insulation
(if present)
Braid Strands Twisted
Under Washer
Keep Wires Flat
Against Case
Optional Plastic or Metal Cable
Clamp (For Physical Strain Relief)
Screw
Split Lock Washer
Flat Washers
External Tooth Lock Washer
Transducer Housing
Figure 3: How To Connect the Cable to a Terminal Block (Example 2)
Содержание Baratron 722B
Страница 1: ...1034138 001 Revision A 5 09 MKS Baratron Type 722B Absolute Pressure Transducer...
Страница 5: ...Table of Contents NW 16 KF Fitting Information 69 Index 71 v...
Страница 6: ......
Страница 8: ...List of Figures and Tables viii...
Страница 58: ...This page intentionally left blank 58...
Страница 62: ...This page intentionally left blank 62...
Страница 70: ...This page intentionally left blank 70...