Chapter Two: Installation
Electrical Information
31
The 622D unit has a 5-pin terminal block connector, whose post assignments are listed in Table 8.
The 626D uses a 15-
pin Type “D” connector
; its pinouts are shown in Table 9. The 623H
transducer uses two (2) terminal blocks: a 5-post block known as TB1, and an 8-post block
designated as TB2. The post assignments for both of these blocks are shown in Tables 10 and 11.
Post Assignments of Type 622D Transducer
Post Number
Assignment
1
Power Common
2
Signal Common
3
Signal Output
4
- 15 VDC
5
+ 15 VDC
Chassis Ground is made by using the screw labeled
CHASSIS
on the endcap
Table 8: Post Assignments of Type 622D Transducer