Interface Cables
Chapter Two: Installation
22
Example 1: Preferred Method To Connect Cable
(shown on a transducer)
Figure 1: Preferred Method To Connect an Overall Metal Braided Shielded Cable
Example 2: Alternate Method To Connect Cable
(shown on a transducer)
Figure 2: Alternate Method To Connect an Overall Metal Braided Shielded Cable
Use this method when cable clamp is not available
Transducer
Overall Insulation
(if present)
Bare Metal Cable Clamp
Making Firm Contact To Braid
Braid Here Is Desirable
(but not usually necessary)
Keep Wires and Braid
Flat Against Case
Metal Cable Clamp
Screw
Split Lock Washer
External Tooth Lock Washer
Transducer Housing
Optional Plastic or Metal Cable
Clamp (For Physical Strain Relief)
Transducer
Overall Insulation
(if present)
Braid Strands Twisted
Under Washer
Keep Wires Flat
Against Case
Optional Plastic or Metal Cable
Clamp (For Physical Strain Relief)
Screw
Split Lock Washer
Flat Washers
External Tooth Lock Washer
Transducer Housing