pg. 7
910 Functions
User Switch
I/O Connector
Red or Green LED Status Indicator
User Switch
The user switch has the following functions:
1. Vacuum Zero adjustment (VAC! Command)
2. Transducer firmware upgrade mode
Refer to pages 17, 18 and 48 for operation procedure.
If the user switch is activated by accident and vacuum Zero or Atmospheric adjustment is executed
the original factory adjustment can be recovered using the FD!VAC command.
(See Factory default page 21).
If the transducer is delivered with customer specified parameters the User Switch is disabled.
For enabling the switch see page 23.
LED Status Indicator
The red/green LED status indicator has the following stages:
1. GREEN
Normal operation
2. 2 sec. RED
Power on sequence
3. GREEN 1 sec. flash cycle
Test mode TST!ON (see page 23)
4. 3 × GREEN flash
User Adjustment executed successfully
5. 3 × RED flash
User Adjustment failed
6. 2 sec. RED
User switch disabled
7. RED
MicroPirani/Piezo sensor failure (see page 20)
8. OFF
Firmware upgrade mode (see page 48) or Power off