How To Configure the Trip Points
Chapter Five: Operation
76
7. Use the Adjust knob to set the normal state for TP A.
The Adjust knob alternates between ENERGIZED and DE-ENERGIZED.
8. Press an arrow key to move the cursor to the TP B entry.
The system responds by moving the cursor to the TP B entry.
9. Use the Adjust knob to set the normal state for TP B.
10. Press the down arrow key ( [
] ) once to display the channel selection screen.
The system responds by displaying the channel selection screen.
TP A Channel
#1
TP B Channel
#1
11. Use the Adjust knob to assign a channel for TP A.
The system responds by scrolling through Channels 1 through 3. Channel 1 is available
on the 670 instrument itself; Channels 2 through 3 are available when the 670 unit is
connected to a 274 multiplexer.
Note
The Adjust knob scrolls through Channels 2 and 3 even when the 670
instrument is not connected to a 274 multiplexer. This enables you to
configure the 670 instrument in one location and then insert it into a
system with a 274 multiplexer, without modification.
12. Press an arrow key to move the cursor to the TP B entry.
The system responds by moving the cursor to the TP B entry.
13. Use the Adjust knob to assign a channel for TP B.
Refer to step 11 for instructions on selecting a trip point channel.
14. Press an arrow key to move to the TP Latch screen.
15. Use the Adjust knob to set the latch for each trip point.
16. Once the trip points are configured, press the [
TRIP POINTS
] key again to return to the
pressure display screen.