Chapter Five: Operation
How To Interface to the Type 274 Multiplexer
69
How To Interface to the Type 274 Multiplexer
The 670 instrument can interface to the Type 274 multiplexer and display the pressure signal
from up to three pressure transducers, individually. Figure 16, page 69, shows the 670
instrument connected to a 274 multiplexer.
Connecting the Cables
1. Disconnect the power cords from the 670 instrument and the 274 multiplexer.
3
2
1
9
6
5
1
11
5
V
On
Off
Power
50 - 60 Hz
75 VA Max
Fuses
180 - 264 V
0.315A (T)
90 - 132 V
0.63A (T)
Voltage
Fuses
Se
ri
a
l I
n
te
rf
a
c
e
I/
O
9
15
9
81
5
6
1
19
37
1
20
Type 670
Type 274
Pressure Transducer
9
5
6
1
9
5
6
1
9
5
6
1
9
6
5
1
15
9
81
15
9
81
ACCESSORY
ELEC
CH 3
CH 2 CH 1
DVM
REMOTE
CB270-2
CB270-2
CB670-1(shown) or CB670-2 (Y-cable)
S
ig
nal
He
ad
4
Figure 16: Connecting to a Type 274 Multiplexer
2. Use a CB670-X cable to connect the I/O connector on the 670 instrument to the DVM
connector on the 274 multiplexer.
The I/O connector is a 37-pin female connector; the DVM connector is a 25-pin female
connector. The CB670-1 cable is a straight cable; the CB670-2 cable is a Y-cable that
allows access to the trip point relays, and the remote functions (zero, range select, and
trip point latching). Refer to Appendix D: Interface Cables for a Type 274 Multiplexer,
page 145, for a description of the cables.
3. Use a CB270-2 cable(s) to connect the pressure transducer(s) to the “CH 1,” “CH 2,” or
“CH 3,” connector(s) on the 274 multiplexer.
4. Use a CB270-2 cable to connect the Head connector on the 670 instrument to the ELEC
connector on the 274 multiplexer.
The Head connector (upper connector on the Signal Conditioner board) is a 15-pin
female connector; the ELEC connector is a 15-pin male connector.