Interface Cables
Chapter Two: Installation
8
Interface Cables
As of January 1, 1996, all products shipped to the European Community must comply with the
EMC Directive 89/336/EEC, which covers radio frequency emissions and immunity tests. MKS
products that meet these requirements are identified by application of the CE Mark.
The 248A valve meets CE Mark requirements, per EMC Directive 89/336/EEC. To ensure
compliance when installed, an overall metal braided shielded cable, grounded at both ends, is
required during use.
Generic Shielded Cable Description
MKS offers a full line of cables for all MKS equipment. Should you choose to manufacture your
own cables, follow the guidelines listed below:
1. The cable must have a
braided
shield, covering all wires. Neither aluminum foil nor
spiral shielding will be as effective; using either may nullify regulatory compliance.
2. The connectors must have a metal case which has direct contact to the cable’s shield on
the whole circumference of the cable. The inductance of a flying lead or wire from the
shield to the connector will seriously degrade the shield’s effectiveness. The shield
should be grounded to the connector before its internal wires exit.
3. With very few exceptions, the connector(s) must make good contact to the device’s case
(ground). “Good contact” is about 0.01 ohms; and the ground should surround all wires.
Contact to ground at just one point may not suffice.
4. For shielded cables with flying leads at one or both ends; it is important at each such end,
to ground the shield
before
the wires exit. Make this ground with absolute minimum
length. Refer to Figures 1 and 2, page 9. (A ¼ inch piece of #22 wire may be
undesirably long since it has approximately 5 nH of inductance, equivalent to 31 ohms at
1000 MHz). After picking up the braid’s ground, keep wires and braid flat against the
case. With very few exceptions, grounded metal covers are not required over terminal
strips. If one is required, it will be stated in the Declaration of Conformity or in the
instruction manual.
5. In selecting the appropriate type and wire size for cables, consider:
A. The voltage ratings;
B. The cumulative I
2
R heating of all the conductors (keep them safely cool);
C. The IR drop of the conductors, so that adequate power or signal voltage gets to the
device;
D. The capacitance and inductance of cables which are handling fast signals, (such as
data lines or stepper motor drive cables); and
E. That some cables may need internal shielding from specific wires to others; please
see the instruction manual for details regarding this matter.
Содержание 248A
Страница 2: ...107428 P1 Rev L 8 06 MKS Type 248A B C Control Valve...
Страница 5: ...Table of Contents iv Index 27...
Страница 7: ...List of Figures and Tables vi...
Страница 11: ...Safety Procedures and Precautions 4 This page intentionally left blank...
Страница 19: ...Setup Chapter Two Installation 12 This page intentionally left blank...
Страница 21: ...General Information Chapter Three Operation 14 This page intentionally left blank...
Страница 27: ...Assembly Chapter Four Maintenance 20 This page intentionally left blank...
Страница 29: ...General Information Chapter Five Repair 22 This page intentionally left blank...
Страница 31: ...General Information Specifications 24 This page intentionally left blank...