31
Left Vacuum
Sequence Timeout
The vacuum setpoint was not reached in the
alloted time. Check for vacuum leaks such as
the doors not sealing. Increase the vacuum
timeout value.
vacuum pump, vacuum
sensor, galil controller,
galil output or analog
card.
Right Vacuum
Sequence Timeout
The vacuum setpoint was not reached in the
alloted time. Check for vacuum leaks such as
the doors not sealing. Increase the vacuum
timeout value.
vacuum pump, vacuum
sensor, galil controller,
or galil analog card.
Environment
Pressure Reading
Error
The value is out of bounds. Check the
connection to the measuring device. Check
the analog output of the measuring device.
pressure transducer,
galil controller, or galil
analog card.
Moisture Level
Reading Error
The value is out of bounds. Check the
connection to the measuring device. Check
the analog output of the measuring device.
moisture monitor or
sensor, galil controller,
or galil analog card.
Oxygen Level
Reading Error
The value is out of bounds. Check the
connection to the measuring device. Check
the analog output of the measuring device.
oxygen monitor or
sensor, galil controller,
or galil analog card.
Left Vacuum
Pressure Reading
Error
The value is out of bounds. Check the
connection to the measuring device. Check
the analog output of the measuring device.
vacuum sensor, galil
controller, galil analog
card.
Right Vacuum
Pressure Reading
Error
The value is out of bounds. Check the
connection to the measuring device. Check
the analog output of the measuring device.
vacuum sensor, galil
controller, or galil analog
card.
Helium Level
Reading Error
The value is out of bounds. Check the
connection to the measuring device. Check
the analog output of the measuring device.
helium analyzer, galil
controller, or galil analog
card.