SECTION 4: OPERATION
31
b. If the remote signal setting is for “monitoring is enabled by remote signal being turned
ON” [S Rte Sgl On]:
(Example of remote signal: Service signal from the water softener
(*)
)
• When the preset time of the monitor interval [S Intvl] has passed, if the remote
signal is ON, the remote signal delay time S Rte Sgl dl[] will start. If the remote
signal is continuously ON while preset remote signal delay time, monitoring will start
soon after the remote signal delay time is finished.
• When the preset time of the monitor interval [S Intvl] has passed, if the remote
signal is OFF, the Colormetry keeps in standby until the remote signal is change
back to ON. When the remote signal is back to ON, the preset remote signal delay
time [S Rte Sgl dl] will start. If the remote signal is continuously ON while preset
remote signal delay time, monitoring will start soon after the remote signal delay
time is finished
Remote signal delay time [S DelayTime]: 10 seconds
ON
OFF
Remote signal
Monitoring
10 seconds
or more
10 seconds
No monitoring
Monitoring interval [S Intvl]: 60 minutes
Monitoring
interval
60 minutes
Monitoring
interval
60 minutes
Monitoring
10 seconds
Monitoring time
Monitoring time
Precautions on operation with remote signal input
If the remote signal turns to disabling monitoring in middle of monitoring process, the
monitoring process will still be carried out to completion. However, the verification of
monitoring result is different by which state of monitoring when the remote signal is
received by the Colormetry
In case the remote signal turns to disable monitoring after water discharging from the
drain-water tube is complete and drainage is stopped by the solenoid valve in the
Colormetry is closed, the monitoring result will remain valid.
In case the remote signal turns before water discharge from the drain-water tube
completed and the solenoid valve is still open, the monitoring result will be processed
as follow:
a. If the concentration is evaluated as being lower than the alarm set point:
The result will remain valid and will be displayed as usual.
b. If the concentration is evaluated as being the alarm set point and up:
The evaluation will be invalidated and the previous result will be displayed.
Monitoring is disabled under the following conditions:
• When the remote signal is turned OFF while the remote signal setting is [S Rte Sgl
On].
• When the remote signal is turned ON while the remote signal setting is [S Rte Sgl Off].