SECTION 4: OPERATION
25
SECTION 4: OPERATION
1. Operation
1.1 Device State Check
When the power of the device is turned on, the device state check starts automatically.
• If the device is working correctly, the check process completes in about 2 to 12 minutes.
(The check process time varies depending on the model, instantaneous flow rate, and
concentration of sample water.)
• At its completion of the check process, the buzzer sounds 4 times and monitoring starts.
z
Remote signal
When the remote signal is utilized, the device keeps in standby until the remote signal permits
monitoring.
z
Abnormal state
If the device is not in a normal state, the check is repeated. During a repeated check, “Self
Check Retry” is displayed.
If the device is still not in a normal state for the repeated check, the buzzer sounds, an error
message is displayed in the LCD, and the device enters the standby mode.
“1. Self-diagnosis Function” in SECTION 6 on page 46
NOTE
“Wash F :F265” or Wash Flow F:F086” may occur and buzzer may sound during first operation
after replacing the fiber filter in spite that the ball valve is open and raw water
(*)
pressure is
applied.
This is an initial phenomenon caused by bubbles in the filter casing assembly. It is not a device
error.
If the same alarm occurs, repeat the manual monitoring several times.
“1.3 Manual Monitoring” in SECTION 4 on page 26