SECTION 4: OPERATION
30
2.2 Colormetry Monitoring Timing
2.2.1 Automatic
Monitoring
The interval between monitoring is set in the monitor interval [S Intvl] setting at the setting
mode [Set Mode] (The interval is selectable in 30minutes increments between the 30 to 240
minutes range).
The initial time point of monitor interval [S Intvl] is the first monitoring which carried out soon
after the Colormetry is supplied the power or is reset. The next monitoring will start after the
preset time of the monitor interval [S Intvl] from this initial time point.
If the monitor interval [S Intvl] setting is changed, the initial time point is also changed to the
monitoring soon before the change is made. And the monitoring will start after the new
setting time from the last monitoring.
[1] Without remote signal input
Monitoring is performed periodically at the monitor interval [S Intvl].
[2] With remote signal input
a. If the remote signal setting is for “monitoring is enabled by remote signal being turned
OFF” [S Rte Sgl Off]:
(Example of remote signal: Regeneration signal from the water softener
(*)
)
• When the preset time of the monitor interval[S Intvl] has passed, if the remote
signal is OFF, the remote signal delay time [S Rte Sgl dl] will start. If the remote
signal is continuously OFF while preset remote signal delay time, monitoring will
start soon after the remote signal delay time is finished.
• When the preset time of the monitor interval [S Intvl] has passed, if the remote
signal is ON, the Colormetry keeps in standby until the remote signal is change
back to OFF. When the remote signal is back to OFF, the preset remote signal
delay time [S Rte Sgl dl] will start. If the remote signal is continuously OFF while
preset remote signal delay time, monitoring will start soon after the remote signal
delay time is finished.
Remote signal delay time [S DelayTime]: 10 seconds
ON
OFF
Remote signal
Monitoring
10 seconds
10 seconds
No monitoring
Monitoring interval [S Intvl]: 60 minutes
Monitoring
interval
60 minutes
Monitoring
interval
60 minutes
Monitoring
10 seconds
Monitoring time
Monitoring time
Water softer
is
regenerating