WARNING:
Install and use this appliance in accordance with the operation manual for safety and EMC (Electromag-
netic Compatibility). If it is not installed and used in accordance with the operation manual, it may cause
interference to other equipment and/or other risk.
To prevent fire or shock hazard, do not expose this appliance to rain or moisture.
This appliance must be earthed.
In Europe, use the AC power cord according to the recommendations as below, in order to comply with
EN60601-1 and EN60950.
Connect to the 230V receptacle of the room or the host equipment.
The AC power cord should be VDE approved and consist of core size 1mm
2
or bigger, length 2.5m or
shorter cord with IEC320/C13 type, 250V 10A or higher rating connector and CEE(7)VII type or IEC 320-
2.2/E type, 250V 10A or higher rating plug.
Use the parallel cable according to the recommendations as below, in order to comply with EN60601-1-2.
The parallel cable with appropriate plug should be 2 m long or shorter, comply with IEEE1284 standard,
shielded, 62
Ω
impedance, ELECOM Co. Ltd. printer cable No. CPC-D2 or equivalent one.
Use the SCSI cable according to the recommendations as below, in order to comply with EN60601-1-2.
The SCSI cable with appropriate plug should be 1 m long or shorter, comply with SCSI II standard,
Mitsubishi Electric Corporation JC-50V50V or equivalent one.
This product is to be employed with medical equipment, just for reference purpose, not for medical
diagnostic purpose.