E-SOUM-OR Rev M
43
System Contents
E-SOGEN-OR
SonicOne O.R console
1 ea.
SOM-HP
SonicOne O.R handpiece
2 ea.
BCM-CW
Counter wrench for SonicOne O.R handpiece
2 ea.
BCM-2W
T-Wrench
2 ea.
BCM-SS
Hard Tissue Probe cover
2 ea.
BCM-H2
Soft Tissue Probe Cover
2 ea.
T
able 11.2 System contents
Components and quantities included with the system may change over time, please check with your Misonix representative
for the most current configuration.
UL (IEC) 60601-1 Classification
Class 1 Equioment
Type B Equipment
Ordinary Equipment
12.
Service, Repair And Technical Correspondence
WARNING 1.4 Proper system grounding can only be ensured when an approved, hospital-grade receptacle and matching
power cord are used. To avoid the risk of electric shock, this equipment must only be connected to a supply
with protective earth. Install plug and receptacles as per local regulations before operating the unit. Power
cord, plug and receptacle should be examined to verify that they are in good working condition before
connecting the console. Never pull on the power cord to remove it from the receptacle.
WARNING 12.1 Connecting the console to a power outlet with inadequate voltage or frequency may cause the unit to
malfunction or to create a shock or fire hazard. Confirm that the voltage selector switch in the console rear
is set to the local voltage setting and ensure that the correct fuses are being used. Refer to section 12 on
fuse replacement
12.1.
Fuse Replacement
Model BCM-GN / E-BC06
Fuse Specifications
Line Voltage
Manufacturer
Manufacturer P/N
Rating
Description
120 VAC, 60 Hz
Cooper/Bussman
GDB-4
250V @ 4 A
Fast Acting, Low Break-
ing
220/230/240 VAC, 50/60 Hz
Littlefuse
021702.5
250V @ 2.5A
Fast Acting
Table 12.1.1 Console fuse specifications model E-SOGEN-OR