9.3.
Sterilizing By Steam Autoclave
Sterilization Methods and terminology are based on ANSI/AAMI ST81 AND EN ISO 17664:2004 standards.
9.3.1
Reusable, Autoclavable Components
BCM-HP
Handpiece
BCM-BW
T-Wrench
BCM-CW
Handpiece/Counter Wrench
BCM-SS
Hard Tissue Probe Cover
BCM-H2
Soft Tissue Probe Cover
9.3.1
Validated Steam Sterilization Cycles
With Sterilization Tray
Wrapped, No Tray
132°C (270°F)
134-137°C
(274-279°F)
134°C/273°F
Configuration
Items wrapped, NO TRAY
3
Items wrapped, NO TRAY
3
Items wrapped, NO TRAY
3
Cycle
Prevacuum
Prevacuum
Gravity
Preconditioning Pulses
4
4
None
Minimum Exposure
4 minutes*
3 minutes*
20 minutes
Minimum Dry Time
45 minutes
30 minutes
5 minutes
Table 9.3.2 Steam sterilization cycles
NOTE 9.1
Allow reusable, autoclavable items to gradually return to room temperature after steam sterilization and
prior to usage.
1
Tray wrapped in Kimberly Clark KC300 KIMGUARD sterilization wrap.
2
Tray wrapped in Kimberly Clark KC300 or KC600 KIMGUARD sterilization wrap.
3
Items wrapped in Kimberly Clark KC300 or KC600 KIMGUARD sterilization wrap.
*Exposure time can be increased up to a maximum of 18 minutes to comply with local requirements and/or recommenda-
tions of the World Health Organization (WHO), Robert Koch Institute (RKI), etc. Misonix Inc reusable medical devices are able
to sustain such sterilization cycles.
132°C (270°F)
134-137°C
(274-279°F)
Configuration
Items placed in Misonix Sterilization
Tray E-SYSTRAY
1
or E-SYSTRAY-2
1
Items placed in Misonix Sterilization
Tray E-SYSTRAY
2
or E-SYSTRAY-2
2
Cycle
Prevacuum
Prevacuum
Preconditioning Pulses
4
4
Minimum Exposure Time
4 minutes*
3 minutes*
Minimum Dry Time
30 minutes
30 minutes