E-SOUM-OR Rev M
28
The console can be placed on an appropriate table or cart outside of the sterile field. Ensure that the pump head on the con-
sole right is installed. Refer to section 12.2 if the pump head is not yet installed.
Figure 7.1.1
The console features air vents on the bottom. When installing the unit, ensure that these vents are not blocked in a way that
would prevent the circulation of air around the unit.
Figure 7.1.2 Underside view of console with air vents
NOTE 7.1
Adequate air circulation is needed to cool electronic components inside of the unit. Do not block the
cooling fan at the console rear or the air vents on the console bottom. Do not place the unit on a towel,
foam or other soft surface since the material may block the air vents. Blocking these vents may cause unit to
overheat and malfunction or create a shock hazard. A clear drape can be used to protect the console front
panel but do not cover the pump housing or other console portions.
Adjust the grip of the V-notches
Evenly adjust the V-notches to their fully
opened position by turning the adjust-
ment wheels.
1 Air vents