![Misonix SonicOne O.R. Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/misonix/sonicone-o-r/sonicone-o-r_instructions-for-use-manual_1806348022.webp)
E-SOUM-OR Rev M
22
1. Menu Button
A-F Custom buttons
Ultrasound Timer
The ultrasound timer records the elapsed time, in which the ultrasound was activated with the footswitch. The timer
can be re-set to zero via the secondary screen.
In the event of error, such as a Mechanical Limit or an Electrical Fault, the main screen is replaced by alert screens. Refer to
section 6.4 for a description of these warnings.
Options Screen
The Options Screen allows the user to do the following; re-set the elapsed ultrasound time, save and choose presets and
adjust the display contrast.
1
A B C D E F
Figure 6.2.2 Options screen
Ultrasound timer
The elapsed ultrasound time can be re-set to 00:00 by pressing either A or B.
Presets
Preferred settings for amplitude, pulse and flow can be saved as two presets. A third preset features the default
settings, which cannot be customized. A sub-screen for presets 1, 2 and 3 can be accessed by pressing either C or D.
Contrast
The display contrast can be adjusted. Press E to increase and F to decrease contrast.
Pressing the menu button toggles from the options to the help screen.