E-SOUM-OR Rev M
32
7.5.
Perform System Check
System Check
Enable Ultrasound
Switch to Enable Mode using enable/standby button. Confirm that Amplitude setting is
FILLED GREEN.
Depress footswitch
Direct ultrasonic tip toward suitable reservoir to collect irrigant. Depress footswitch.
Confirm Function
Console emits a bell chime.
Irrigant will be pumped from console towards handpiece.
Ultrasonic tip emits buzzing sound and irrigant exits tip as fine spray.
Ultrasound timer counts up in 1-second increments.
Release footswitch
Release footswitch.
Ultrasound and Flow output stop.
Ultrasound timer freezes at last reading.
Function Confirmed
Reset ultrasound timer as desired.
System is now ready for use.
Function NOT confirmed
Console alerts of Mechanical Limit or Electrical Fault or does not respond as expected.
Refer to troubleshooting section for next steps.
Table 7.5 System check
CAUTION 7.8
The system check should always be done in advance of preparing patient for surgery to minimize risk to
patient in case of system malfunction.
The SonicOne O.R system is now ready for use. Refer to section 1.0 for general safety statements, indications and adverse
affects and section 5.0 for use of main system functions.
8.
Handpiece Assembly And Disassembly
The SonicOne
O.R system can accommodate different tip designs. The tip choice is determined by the surgeon’s preference.
8.1.
Handpiece Assembly
Perform an inspection of handpiece and all components prior assembly.
Handpiece Inspection
Inspect Handpiece
Inspect the black handpiece housing for any visual cracks. Inspect the front metallic portion probe
for surface damage like nicks, gouges and cracks. Replace if damaged.
Inspect Mating Surface
Inspect mating face of handpiece to verify that it is clean and dry.
Table 8.1 Handpiece inspection