Spirolab
Rev.1.8
Page 9 of 60
1.2.2
Turbine
Disposable turbine
WARNING
If you decide to use the spirometer with a “disposable”
turbine, it is important to use a new turbine for each
patient.
Adequate hygiene, safety and operational conditions of
“disposable” turbines are only guaranteed if said
turbines are preserved in the original, sealed packaging.
“Disposable” turbines are made of plastic; to dispose of
the turbines, please comply with local legal
requirements.
Reusable turbine
WARNING
The correct use of a “
reusable
” turbine is guaranteed
only and exclusively if it is “clean” and free from foreign
bodies that alter the motion of the blades. If a reusable
turbine is not properly cleaned, it can cause cross
infection between patients. Periodical cleaning of the
device is only sufficient if that device is used exclusively
by the same patient. For further details about cleaning,
refer to the appropriate section in this User Manual.
The following information is valid for both types of turbine.
Never expose turbine sensors to jets of water or air or allow contact with hot liquids.
To avoid malfunctions or damage, do not introduce dust or foreign bodies into the turbine
sensor. The presence of foreign bodies (such as hairs, saliva etc.) inside the turbine flow
meter can compromise the accuracy of readings.
1.2.3
Mouthpiece
The single-use mouthpieces that come supplied with the device are provided as a sample
to demonstrate the correct type and size. They must be considered clean but not sterile.
We recommend that you contact the local distributor where you bought your spirometer
to obtain additional mouthpieces. These are usually made of cardboard or plastic and are
always for single use only.
WARNING
Use biocompatible mouthpieces to avoid problems for the patient; unsuitable
materials might cause the device to malfunction and compromise the accuracy of
readings.