Conformed standard
This instrument has conformed to the following standard.
Annex D (Performance level H) of JIS B 7611-2
:
2009
[Non-automatic weighing instruments
-
Metrological and technical requirements and tests
-
Part 2: Measuring instruments used in transaction or certification]
Caution
Please observe the following conditions strictly when this instrument comply the above mentioned standard.
(1) Shield processing
Cables other than power cable must use all shielded cables.
Refer to relative notes, about method of shield processing.
(2) Function setting
As for the details of the function of the value of C Function and Function,
please refer the paragraph 5-2 and 7-2.
CF-03 [Condition of over display] shall be applied with the value of 2.
CF-11 [Effective range of zero set] shall be applied with the value of 0.
CF-13 [Data width of zero tracking] shall be applied with less than 4% of maximum weighing capacity.
F-01 [Digital filter setting] shall be applied with more than the value of 2.
F-05 [Stabilized filter setting] shall be applied with more than the value of 4.
F-06 [Stabilized filter data width] shall be applied with less than the set value of 005.
F-07 [Stabilized filter time width] shall be applied with the set value of more than 01.
F-10 [Detection of stability data range] shall be applied with less than the set value of 4.
F-11 [Detection of stability time range] shall be applied with more than the set value of 2.