4. Notes on installation and connection
1
Installation place
●
In case of connection with the instrument, keep strictly to the following items. If
neglected, it may cause an unexpected failure or a damage to the instrument.
D
Please choose the place without a mechanical vibration and a corrosive gas as the place
where this instrument is installed.
Moreover, be careful not to give unnecessary external force to this instrument.
D
Please avoid use in the environment that moisture adheres to this instrument
Moreover, please never pour water directly.
2
Connection
D
The connecting cable with the torque transducer must use the cable attached to the torque
transducer.
D
Please separate the cable connected with this instrument as long as possible from the noise
sourse, like a power supply line (motor and inverter) or I/O line for control, etc.
It causes the accuracy decrease (zero change and output voltage change, etc.) by the
influence of the noise when wiring near these lines.
Especially, please separate the cable which connects the torque transducer with this
instrument from these lines by 50 cm or more.