10
5
. Analog output
5
−
1. Voltage output
1
Output
Output
±
10 V at the time of the rated output of torque transducer.
The analog output range can be changed by attenuator switch (DIP401).
DIP401
Output range
only 1 is ON
0
±
10 V
only 2 is ON
0
±
5 V
only 3 is ON
0
±
2 V
only 4 is ON
0
±
1 V
Please make the setting of switch as only one.
2
Electrical condition
D
Load resistance
:
2 k
Ω
or more
D
Capacitance load
:
0.1
μ
F or less
5
−
2. Current output (option)
Adjustment can be made when the optoinal current output is set.
Please execute the setting after turning off the power.
1
Output
The following setting of the current output can be made at the rated output from the torque
transducer.
i) When the output of 12.00 mA
±
8.00 mA at 0
〜±
rated output of torque
The current output adjustment switch (SW402) is set to the upper part.
ii) When the output of 4.00 mA
〜
20.00 mA at 0
〜+
rated output of torque.
The current output adjustment switch (SW402) is set to the lower part.
The main setting of the voltage output and the current output is as follows;
a ) When 12 mA of the current output at 0 V of the voltage output
Voltage output
Current output
SW402
GAIN
−
10 V
〜
10 V
4 mA
〜
20 mA
Upper side
1
−
5 V
〜
5 V
4 mA
〜
20 mA
Upper side
2
−
2 V
〜
2 V
4 mA
〜
20 mA
Upper side
5
−
1 V
〜
1 V
4 mA
〜
20 mA
Upper side
8
b ) When 4 mA of the current output at 0 V of the voltage output
Voltage output
Current output
SW402
GAIN
0 V
〜
10 V
4 mA
〜
20 mA
Lower side
2
0 V
〜
5 V
4 mA
〜
20 mA
Lower side
4
0 V
〜
2 V
4 mA
〜
20 mA
Lower side
8
0 V
〜
1 V
4 mA
〜
20 mA
Lower side
F
2
Electrical condition
D
Load resistance
:
510
Ω
or more
Содержание CSA-562B
Страница 1: ...Transmitter for Torque transducer CSA 562B Instruction Manual EN294 1170 F ...
Страница 2: ......
Страница 10: ......
Страница 32: ......
Страница 33: ......