When checking voltage between A-C, set the connecting range to DC・V for the measuring instrument such as digital multimeters
When checking voltage between B-D, set the connecting range to DC・mV for the measuring instrument such as digital multimeters.
Remove the cable for strain gage applied transducer from the terminal board.
①
Connect the cable for strain gage applied transducer to the terminal board again.
(Refer to the paragraph 3)
②
By checking the voltage between the terminal B and D on the terminal board, apply load on the
strain gage applied transducer.
The voltage between B and D is
vary
Check the various connecting condition
according to the paragraph 3.
There is no change
Start Measurement.
Inform us about the contents of failure
and situation at site in details.
Inform us about the contents of failure
and situation at site in details.
3
YES
NO
NO
There may have fears that the strain gage
applied transducer connected has broken or
signal route (signal wire) has cut off.
YES
NO
YES
The voltage between
A-C is DC10 V