19
When checking voltage between A-C, set the connecting range to DC・V for the measuring instrument such as testers
When checking voltage between B-D, set the connecting range to DC・mV for the measuring instrument such as testers.
Remove the cable for strain gage applied transducer from terminal board.
The voltage between A-C is
DC10 V and stable.
Short circuit the B, D and V (-) of the terminals.
Output voltage is
stable.
Check the various connecting condition according to the paragraph 2.
Especially, check any effects from the noise source such as an
inverter or like that.
There is no change
Start Measurement.
Inform us about the contents of failure
and situation at site in details.
Inform us about the contents of failure and
situation at site in details.
4
YES
NO
NO
YES
NO
YES