![Minebea Intec Signum 3 Ex Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/minebea-intec/signum-3-ex/signum-3-ex_operating-manual_1801214054.webp)
54
Signum 3 Ex
Averaging (Animal Weighing) Application
V
Averaging (Animal
Weighing) Application
V
With the Averaging application, you
can use your weighing platform for
calculating weights as the average
of a number of individual weighing
operations� These individual operations
are also known as “subweighing
operations�”
This function is used to determine
weights under unstable ambient
conditions or for weighing unstable
samples (such as live animals)�
Characteristics
– Averaging started manually or auto-
matically� Configured in Setup under:
appl1
:
anim.wg
:
3.18.
With manual start selected, the
averaging routine begins when
you press a key (provided the start
conditions are met)� With automatic
start selected, averaging begins when
you place the first load on the platform
(provided the start conditions are met)�
– Enter the number of subweighing
operations using the keypad
– Info mode
– Toggle the display from “result of last
measurement” to “current weight” by
pressing the
w
key
– Autmatic printout configured in Setup
under:
Appl1
:
anim.wg
:
3.20.
– Automatic taring of container weight�
Configured in Setup under:
appl1
:
anim.wg
:
3.7.
– Automatic start of averaging when
the scale is turned on and a sample
placed on the platform (provided start
conditions are met)� Configured in
Setup under:
Appl
:
a.start
:
3.8.
– Exit application, delete parameters:
The number of measurements remain in
memory until it is deleted, overwritten
or the application is changed using the
c
key�
The number of measurements also
remains saved after the scale is turned
off�
You can assign different functions to
the
c
key for deleting applications�
When you clear applications, you can
delete either the data stored for all
applications or just selected data stored
for the active application� Configured in
Setup under:
Appl
:
sel.cf
:
3.24.
(Sel� CF function key
c
in
applications)
– Tare function:
If you store a tare (weight value) by
pressing the
)
key, you can later
enter a value manually� The tare value
you enter is added to the stored tare
value�
Setting: menu code 3�25�1 (factory
setting)
A tare value entered manually
overwrites a stored tare value (weight
value)� If you enter a tare value
manually, a tare value (weight value)
stored later overwrites the manually
entered value� Setting: menu code
3�25�2
Configured in Setup under: Signum
®
2
Appl
:
tare. f
:
3.25.
– Restore factory default settings�
Configured in Setup under:
Appl
:
DEF.App
:
9.1.
A number of subweighing operations
are required to form the basis for
calculation of an average weight�
You can enter the desired number
of subweighing operations using the
keypad�
The number you enter is active until
it is overwritten by another number�
It also remains in memory when you
switch to a different application
program, or turn off the scale�
There are three ways to start the
averaging routine:
– Manual start with preset number of
sub-weighing operations:
Place the sample on the platform and
press the
O
key
– Manual start with user-defined number
of subweighing operations:
Place the sample on the platform
and enter the number of weighing
operations using the keypad� Press the
r
key to save the number entered
and begin weighing
– Automatic start with preset number of
subweighing operations:
Measurement begins when you place
the first sample on the platform,
provided the start conditions are met�
Averaging (Animal Weighing) Application
V
Содержание Signum 3 Ex
Страница 97: ...97 Signum 3 Ex 97 Declaration of Conformity...
Страница 98: ...98 Signum 3 Ex Declaration of Conformity...
Страница 99: ...99 Signum 3 Ex 99 Certificates 2017 11 23...
Страница 100: ...100 Signum 3 Ex Certificates...
Страница 101: ...101 Signum 3 Ex Certificates...
Страница 102: ...102 Signum 3 Ex Certificates...
Страница 104: ...104 Signum 3 Ex Certificates...
Страница 113: ...113 Signum 3 Ex...
Страница 117: ...117 Signum 3 Ex Verification of intrinsic safety...
Страница 118: ...118 Signum 3 Ex Verification of intrinsic safety...
Страница 119: ...119 Signum 3 Ex Verification of intrinsic safety...
Страница 120: ...120 Signum 3 Ex Verification of intrinsic safety...
Страница 121: ...121 Signum 3 Ex Verification of intrinsic safety...
Страница 122: ...122 Signum 3 Ex Verification of intrinsic safety...
Страница 123: ...123 Signum 3 Ex Verification of intrinsic safety...
Страница 124: ...124 Signum 3 Ex 124...
Страница 134: ...134 Signum 3 Ex...
Страница 135: ...135 Signum 3 Ex...