8 Troubleshooting
8.1 General Notes
The following hints will enable a technician to do an initial diagnostic or help in case of
incorrect or non-reproducible weighing results after commissioning and calibration.
8.2 Visual inspection
Component
Possible errors
Weighing object
Are all pipes, hoses and cables free from shunt forces?
Are the connections pliable and connected horizontally?
Are elements with a solid connection to the scale in direct contact with the surroun-
dings?
Has friction developed between the weighing object and its surroundings (e.g. dus-
ty openings, …)?
Cable junction box
Has moisture intruded?
Do all soldering and screw connections have secure contact?
Connecting cables
Is the sheath damaged?
Has moisture intruded?
Mounting kit
Has dust collected between the load cell and the overload fuse?
Are the constrainers stuck?
Load cell
Is the load cell mounted in accordance with the load direction (arrow on the face of
the load cell)?
Are the load cells, speciically the bellows, damaged?
Is the sheath of the load cell cable damaged?
Has moisture penetrated into the load cell cable?
8.3 Metrological controls
8.3.1
Checking the zero output signal of the load cell
-
Unload load cell.
-
Disconnect the load cell measuring outputs.
-
Check whether the output voltage without load is within the limits.
Type
Output voltage
D1, C3
0 ±0.02 mV/V
8.3.2
Checking the strain gauge bridge of the load cell
-
Do not exceed the test voltage.
-
Check whether the values of the resistors are within the permissible limits.
8 Troubleshooting
Beam Type Load Cell PR 6207
Minebea Intec
EN-21
Содержание PR 6207
Страница 21: ...6 Connection Beam Type Load Cell PR 6207 Minebea Intec EN 19 ...
Страница 29: ......
Страница 30: ......
Страница 31: ...1 6 ...
Страница 32: ...2 6 ...
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......