6 Connection
6.1 General information
NOTICE
The load cell cable end is protected against contamination and moisture by a
protective sheathing.
The protective sheathing should only be removed shortly before connecting!
-
Protect the cable ends against contamination. Moisture must not get into the open
end of the cable.
-
Do not shorten the load cell connecting cable. Connect the prepared cable end and
roll up the remaining cable.
-
The cable screen may be connected only to the connecting terminals of the indicator.
-
Keep the load cell cables away from power cables.
-
The distance between measurement cables and power cables and/or components
under high voltage should be at least 1 m (reference value).
-
We recommend laying the load cell cables in separate cable trays or armored steel
pipes.
-
Power cables should be crossed at right angles while taking into account the
minimum distance of 1 m (reference value).
6.2 Load cell
Color code
bk
=
black
bu
=
blue
gn
=
green
gy
=
gray
rd
=
red
wh
=
white
ye
=
yellow
bu =
+ supply/LC in
+ supply v load cell input
gn =
+ sense
+ sense
wh = + meas./LC out
+ measuring v load cell output
bk =
- supply/LC in
- supply v load cell input
gy =
- sense
- sense
rd =
- meas./LC out
- measuring voltage/- load cell output
ye =
screen
Screen
6 Connection
Beam Type Load Cell PR 6207
Minebea Intec
EN-17
Содержание PR 6207
Страница 21: ...6 Connection Beam Type Load Cell PR 6207 Minebea Intec EN 19 ...
Страница 29: ......
Страница 30: ......
Страница 31: ...1 6 ...
Страница 32: ...2 6 ...
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......