background image

  

t

   To change the calibration/adjustment unit, select the unit using the “

Q

“ or “

q

“ 

soft key and confirm using the “

l

“ soft key. 

t

  Use the “

o

“ soft key to go back to the “

Internal

“ menu.

t

  Use the “

Q

“ soft key to select the “

ADC conf

i

guration

“ menu item.

  

t

   Use the “

O

“ and “

q

“ soft key to select the “

Save conf

i

guration 

data

“ menu item.

  

t

   To save the configuration, use the “

Q

“ soft key to select “

Yes

“ and confirm 

using the “

l

“ soft key.

y

   The message “

Function activated

“ appears briefly.  

The program then returns automatically to the regular weighing mode.

    To not save the configuration:

t

  Press the “

o

“ soft key to exit the menu. 

y

  The program returns to the next higher menu level.

    To not save data: 

t

  Press the “o“ soft key. The program returns to the 

WP1:Internal

.

t

   Slide the menu access switch to the right (= “closed“ position) and reattach the 

cap.

y

  The device is now in normal weighing mode.

 

h

  Once ADC configuration has been completed, an adjustment of the weighing 

platform (calibration/adjustment and linearization) must be carried out (see 

“Calibration/Adjustment without Weights“ and “External Linearization“).

30

 

Operating Instructions Combics Complete Scales

 

Configuring Weighing Platforms

Содержание CAW3P

Страница 1: ...98648 018 49 Operating Instructions Minebea Intec Combics Series Complete Scales Models CAW3P CAW3S CAS3E CAS3G CAS SPC ...

Страница 2: ...zation 47 Setting the Preload 50 Clearing the Preload 51 Operating Design 52 Weighing Operating Design 52 Menu Operating Design 58 Configuration 59 Setup Overview Parameters 63 Operation 76 Weighing 76 Calibration Adjustment 86 Internal Adjustment for CAS3 Models 90 SQmin Function 91 Data ID Codes 94 Data Interfaces 96 Configuring the Data Interface as a COM Port 99 Data Input Format 101 Data Outp...

Страница 3: ...or various types of dangers These symbols are explained in more detail in Section Safety Instructions h This symbol indicates useful information and tips e 1 This and similar symbols mean that the respective key should be pressed T T This means that this key must be pressed more than once t Indicates a required action y Describes the result of an action 1 If a procedure has multiple steps 2 the st...

Страница 4: ... under ambient conditions requiring higher safety standards you must comply with the provisions as specified in the applicable regulations for installation in your country 3 The operator shall be responsible for any modifications to the equipment and for any connections of cables or equipment not supplied by Minebea Intec and must check and if necessary correct these modifications and connections ...

Страница 5: ...64ED S CASLE SPC 06HCE models platform IP43 Some models IP43 indicator IP69K CASLE SPC models platform IP65 indicator IP44 CAS3G models platform IP67 indicator IP69K CAWxS models are rated to IP69K with the I69 option Complete scales with secured protective caps must be installed and tested by a certified technician If you install an interface port or battery connection after setting up your indic...

Страница 6: ...al Formulation Automatic initialization when the scale is switched on Fast response times Automatic taring when a load is placed on the weighing platform Designation of weight values with up to 4 lines of alphanumeric text Can be controlled via two external computers using various protocols Barcode scanner connection option for entering tare value or IDs 6 units Possibility to input tare values vi...

Страница 7: ... Ethernet TCP IP ModBus TCP 12 RS 232 interface COM1 e g for PC connection Printer connection Digital input Stop light output 13 PS2 e g for Keyboard connection Barcode scanner connection 14 Power cord with country specific plug 15 Input for menu access switch standard or legal for trade mode for WP 1 16 Weighing platform WP 1 connection for analog scales optional for xBPI SBI scales 17 Weighing p...

Страница 8: ...ssarilyextremetemperatures moisture shocks blows or vibration Permissible storage temperature 10 C to 40 C Installation Location Avoid adverse influences at the place of installation Extreme temperatures Aggressive chemical vapors Extreme moisture according to IP protection class Unpacking For devices with a platform size of 60 x 80 cm or larger t Protective gear must be worn safety shoes and if r...

Страница 9: ...Allen wrench key to remove screw no 1 and loosen screw no 2 Turn the red color coded transport locking device by 90 in the direction of the arrow and refasten the transport locking device t Make sure to retighten all screws that belong to the transport locking devices t Place the load plate on the platform 3 The transport lock must be re installed before transporting the weighing platform Weighing...

Страница 10: ...foot t Place the weighing pan on the scale Operating Limits You should not exceed the highest load for weighing platforms The highest capacity for the weighing platform is as follows depending on the load used center side one sided corner load Platform size Center Side Corner 320 240 50 35 20 400 300 130 85 45 500 400 300 200 100 500 400 P 600 400 200 650 500 S 450 300 150 800 600 P 1200 800 400 8...

Страница 11: ... the weighing pan When setting up systems in hazardous areas Zone 2 or 22 any relevant specifications should be observed e g EN60079 14 The design should ensure that moving parts do not cause electrostatic discharges e g roller conveyors Preload Range Zero Set Range The weight of the superstructures that are attached to the weighing platform is designated as a preload The preload is electronically...

Страница 12: ...has received specialized training from Minebea Intec Any installation work that does not conform to the instructions in this manual results in forfeiture of all claims under the manufacturer s warranty 3 Disconnecttheequipmentfromthepowersupplybeforestartingconnectionwork t Place cables from peripheral devices next to the indicator CAW3S CAS3 IP69K t Opening the Combics indicator Loosen the ten ca...

Страница 13: ...Insert all cable wires through the ferrite case wind them around the ferrite case and then reinsert back through the ferrite case t Screw the wires tightly into the clamps See the following pages for terminal pin allocation t Refer to the data sheet or operating instructions of the weighingplatformfordetailsontheassignmentofwire colors signals Ensure any lines that are not assigned are insulated c...

Страница 14: ...utput heavier 34 TXD_COM2 10 LINE_OUT 20 Control output set 35 GND 36 GND A8 terminal assignments 1 EXC Bridge supply voltage 2 SENSE Sense for bridge supply voltage 3 OUT Measuring voltage positive 4 OUT Measuring voltage negative 5 SENSE Sense for bridge supply voltage 6 EXC Bridge supply voltage Keys LED Display A Alibi memory A6 7 Tastatur LED Display PS2 COM1 COM2 Interface PCB for RS 232 485...

Страница 15: ...d cable clamp assembly and shield plate Amp type 826 985 1C and fastening screws Amp type 164868 1 COM1 pin assignments Pin 1 Shield Pin 2 Data output TxD Pin 3 Data input RxD Pin 4 GNO Pin 5 Clear to send CTS Pin 6 Not assigned Pin 7 Internal ground GND Pin 8 Internal ground GND Pin 9 Not assigned Pin 10 Not assigned Pin 11 12V for printer Pin 12 RES_OUT Pin 13 5 V switch Pin 14 Internal ground G...

Страница 16: ...cable from the indicator to an RS 232 PC interface COM1 Indicator side PC side Model type CAW3P CASLE SPC DSUB connector 25 pin D Sub male connector 9 pin or 25 pin 1 Sgn GND 7 5 GND 7 GND TxD 2 2 RxD 3 RxD RxD 3 3 TxD 2 TxD DTR 20 8 CTS 5 CTS CTS 5 4 DTR 20 DTR Model type CAW3S CAS3 Open cable end DSUB connector 9 pin or 25 pin Sgn GND 15 5 GND 7 GND TxD 14 2 RxD 3 RxD RxD 13 3 TxD 2 TxD DTR 12 8...

Страница 17: ...ace connector 9 pin D Submini DB9 with shielded cable clamp assembly and shield plate and fas tening screws Amp type 164868 1 Pin assignment Pin 1 5 V out Pin 2 Data output TxD Pin 3 Data input RxD Pin 4 Clear to Send CTS Pin 5 Internal ground GND Pin 6 Reset Pin 7 Not assigned Pin 8 Data Terminal Ready DTR Pin 9 Load Printer 6 5 1 9 Getting Started Operating Instructions Combics Complete Scales 1...

Страница 18: ...cable from the indicator to an RS 232 PC interface COM1 Indicator side PC side Model type CAW3P CASLE SPC DSUB connector 25 pin D Sub male connector 9 pin or 25 pin 1 Sgn GND 7 5 GND 7 GND TxD 2 2 RxD 3 RxD RxD 3 3 TxD 2 TxD DTR 20 8 CTS 5 CTS CTS 5 4 DTR 20 DTR Model type CAW3S CAS3 Open cable end DSUB connector 9 pin or 25 pin Sgn GND 15 5 GND 7 GND TxD 14 2 RxD 3 RxD RxD 13 3 TxD 2 TxD DTR 12 8...

Страница 19: ...M8 on page 14 NOTE This equipment has been tested and found to comply with the limits pur suant to part 15 of FCC Rules These limits are designed to provide reasonable protection against harmful interference This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with these instructions may cause harmful interference to radio communications ...

Страница 20: ...nstalled wall outlet which has a protective grounding conductor PE Safety Precautions 3 If you use an electrical outlet that does not have a protective grounding conductor ensure that an equivalent protective conductor is installed by a certified electrician as specified in the applicable regulations for installation in your country The protective effect must not be negated by using an extension c...

Страница 21: ... General User Password and press M to confirm y The device in now is Service mode An S appears in the top right hand corner of the display t Press the q soft key several times to select the Device parameters line t Press the O soft key y The Device submenu will open t Select and open the corresponding submenu Repeat this procedure until the desired menu item in the lowest menu level can be opened ...

Страница 22: ...ration Adjustment CAL key function o Ext cal adj factory def wt Ext cal adjust user defined weight Ext lineariz factory def wts Ext lineariz user def wts Set preload 2 Delete preload 2 Key blocked Cal adj sequence Cal then auto adjust o Cal then manual adjust isoCAL function 3 o Off Adjustment prompt Activate external adjustment 2 o Activated Deactivated Parameter for external weight Cal adj weigh...

Страница 23: ...view Parameters section RS 232 1 similar to IS 485 menu for WP1 RS 485 1 similar to IS 485 menu for WP1 o Off COM 1 similar to WP 1 COM 2 similar to WP 1 COM 1 see the Setup Overview Parameters section COM 2 see the Setup Overview Parameters section Control I O ports Terminal data see the Setup Overview Parameters section SQmin SQmin input SQmin WP1 SQmin WP1 0 000 kg SQmin WP2 SQmin WP2 0 000 kg ...

Страница 24: ...e 3 Max cap Available units User definable o Grams g Kilograms kg Carats ct Save parameters Yes oNo oVerifiable Accuracy Class III IIII configuration class Ranges Single range mode D E Min capacity Max cap Multi interval mode D E Min capacity Range 1 Range 2 Range 3 Max cap Multi interval mode D E Min capacity Range 1 Range 2 Range 3 Max cap Available units User definable o oGrams g oKilograms kg ...

Страница 25: ...fiable for use in legal metrology weighing platform Standard configuration Standard Verifiable configuration Verifiable Select using the Q or q soft key Press the O soft key to confirm the setting and open the Configuration menu Accuracy Class This menu item is not shown when the Standard configuration is active When the Verifiable configuration is active for weighing platforms verified or verifia...

Страница 26: ...ax capacity 15 000 kg smallest scale interval d 0 005 kg yields 3000 scale intervals In legal metrology the total number of intervals must be no more than 3000 e and when using multi interval scales there must not be more than 3000 e intervals per range In standard operation as opposed to legal metrology you can define a Super Range weighing instrument of over 3000 intervals These parameters howev...

Страница 27: ...configuration data record using the Q or q soft key Standard or Verifiable verifiable configuration The default setting depends on the data record t Open the menu for configuring A D converter parameters In this example the menu for ADC configuration of a weighing platform for use in legal metrology is opened If Standard was selected previously then the Accuracy tass is not displayed t Open the fi...

Страница 28: ...el y This will apply all parameters t Press M or oo to exit the Setup menu 0 0 0 2 In the example shown here a single range scale in Verifiable configuration with a maximum capacity of 6 000 kg is modified the verification scale interval e is changed from 0 001 kg to 0 002 kg in accordance with the maximum permitted value of 3000 verification scale intervals Press the l soft key to confirm the cha...

Страница 29: ...ts menu item This menu lets you enable or disable the weight units available under Weight unit 1 Weight unit 2 and Weight unit 3 t Select the respective unit using the Q or q soft key and confirm using the l soft key In most cases you will not need to change defined values Available weight units are marked by a The weight unit used for configuration of weighing ranges cannot be blocked t Use the o...

Страница 30: ...returns automatically to the regular weighing mode To not save the configuration t Press the o soft key to exit the menu y The program returns to the next higher menu level To not save data t Press the o soft key The program returns to the WP1 Internal t Slide the menu access switch to the right closed position and reattach the cap y The device is now in normal weighing mode h Once ADC configurati...

Страница 31: ... for multiple range scales ADC Configuration with Load Cell s Connected Procedure 1 Open the menu access switch see Analog Digital Converter ADC Configuration 2 Activate the Service mode see Service Mode 3 Configure WP 1 see Analog Digital Converter ADC Configuration 4 Set single range mode for example see Analog Digital Converter ADC Configuration 5 Select the units see Analog Digital Converter A...

Страница 32: ...810 m s2 On delivery the geographical data for Germany Zone D are entered in the output device It is recommended to use the geographical data settings for Germany Zone D when adjusting and delivering the weighing equipment within Germany Entering exact geographical data will lead to a higher level of accuracy but will also restrict the tolerance zone Setup Information It is only possible to enter ...

Страница 33: ... parameters Entering Geographical Latitude and Elevation t Use the Q or q soft key to select the corresponding input field t Enter the number via the keypad and confirm using the l soft key y The next input field is selected In this example the geographical data are entered for the respective platform as a value pair Latitude and Altitude After this data was saved and the scale returned to weighin...

Страница 34: ...e menu access switch to the right closed position and reattach the cap y The display goes out and the device restarts Then weighing mode is active Adjusting Scales in Operating Mode See also Calibration and Adjustment in the chapter called Operation t Open the Device parameters menu for the respective weighing platform e g WP 1 INTERNAL Open the Calibration adjustment submenu CAL key function menu...

Страница 35: ... press to cancel the adjustment Latitud C EXT DEF y The display Latitud appears briefly y The geographical latitude of the place of installation is shown in degrees north or degrees south here the latitude setting for Germany Zone D t Press J to confirm the displayed value or press to cancel the adjustment y You are prompted to place the required weight on the platform e g 5 0 kg The subsequent st...

Страница 36: ...e described in the chapter entitled Operation under Calibration and Adjustment t Slide the menu access switch to the right closed position and reattach the cap y The display goes out and the device restarts Then weighing mode is active If adjustment is carried out using a verifiable configuration data record the lines for display of metrological data lines 1 and 2 show the data valid for use in le...

Страница 37: ...Lin wt 4 can however only be changed in the Service mode y The current values for the user defined calibration weight and the 4 linearization weights are displayed In this example the value for the external user defined adjustment weight is changed to 6 000 kg t Press 6 000 and confirm with the l soft key y The Lin wt 1 input field is selected In this example the value for linearization weight 1 i...

Страница 38: ...ters is displayed t Use the Q or q soft key to select and open the Calibration adjustment menu using O t Use the O soft key to open the Cal adj sequence menu t Use the Q or q soft key to select the Cal then manual adj menu factory setting and confirm with the l soft key t Use the o soft key to go to the next menu level t Use the O soft key to open the Activate ext adj menu t Use the Q or q soft ke...

Страница 39: ...he scale is now in Service mode without this being displayed t Trigger the previously set function using the J key C EXT D C EXT DEF 12 y The display shows S_CAl indicating that the scale is in Service mode t The corresponding menu item can be selected using the Select soft key and carried out using the J key t If you cancel the function using the key or by restarting the scale using the e key you...

Страница 40: ...the allocation of the J Key for Calibration Adjustment andtheadjustmentweightconsistsofseveralpieces then they should be placed on the platform in series at short intervals When the weighing instrument has stabilized the weight on the scale is accepted as the calibration weight after a predefined interval and the weighing instrument is calibrated adjusted with this weight The difference since the ...

Страница 41: ... Adjust without weights The zero point is only displayed for a verified scale with d e External Calibration Adjustment with a User Defined Weight Configuration Select the CAL key function Ext cal adj user def wt menu item external adjustment using a user defined weight Note The menu items Ext cal adj factory def wt external adjustment with a factory defined standard weight Ext cal adj user defined...

Страница 42: ...t refer to the note under External Calibration Adjustment with Factory Defined Weight Default Weight y After a brief pause the difference since the last span adjustment is displayed calibration t If you only want to perform a calibration press the key to cancel the calibration adjustment procedure t Press J to adjust the scale y At the conclusion of the calibration procedure the calibration weight...

Страница 43: ...l as via the expansion PCBs for WP 1 or WP 2 This function is also accessible without activating the Service mode Adjustment Without Weights Purpose In the Service menu adjustment without weights can be carried out by entering the characteristic data of the load cells 3 Adjustment without weights may not be carried out on weighing equipment used in legal metrology Setup information Adjustment with...

Страница 44: ...Calibration adjustment menu using O t Use the Q or q soft key to select and use O to open the Adjust without weights menu t Use the O soft key to open the Input parameters menu y The Input menu is displayed Enter the nominal capacity and resolution of the load cells in kg and the sensitivity of the load cells in mV V in the corresponding input fields The maximum capacity is usually less than the v...

Страница 45: ...as an average value If an average value is entered for all cells or if fewer than 4 load cells are present enter 0 in the remaining fields Range of permitted values 0 01 5 mV V t Press 1 944 e g to enter a sensitivity of 1 944 mV V t Confirm using the l soft key y The Sensitivity 2 input field is selected t Either enter a value or set all remaining input fields to 0 t Confirm each using the l soft...

Страница 46: ...ustment menu using O y The CAL key function submenu is displayed t Use the Q or q soft key to select the corresponding menu item and confirm with l y The menu item is marked by a circle o Note The functions that can be configured in the CAL key function submenu depends on the selected weighing platform and its configuration data Functions that cannot be accessed are not displayed in the selection ...

Страница 47: ...splay of geographical data has no effect on this function 3 Once linearization has been completed the J key must be reallocated back to its original function in the Setup menu e g external calibration adjustment with default weights Procedure t For scales used in legal metrology slide the menu access switch to the left open position y The display goes out and the device restarts Then weighing mode...

Страница 48: ...ult weights Configuration t Set the linearization weights see Entering Adjustment and Linearization Weights Procedure t For scales used in legal metrology slide the menu access switch to the left open position y The display goes out and the device restarts Then weighing mode is active t Press to unload and zero the scale t Press J to start the external linearization y The display L EXT U appears b...

Страница 49: ...ove all linearization weights from the weighing platform y The zero point is applied automatically after a brief time The indicator automatically switches to weighing operation When linearization is complete the GMP compliant printout shown on the left is generated t Slide the menu access switch to the right closed position 24 02 2016 10 15 Model CAW3E Ser no 12345678 Vers H0 111 031115 BVers 01 6...

Страница 50: ...Intec logo is displayed briefly after which the device is in normal weighing mode t Press to unload and zero the scale y Display after the scale has been zeroed t Place the preload weight on the weighing platform t Press J to start Set preload y The display SET PREL appears briefly After a short period of time the preload will be applied and the indicator will automatically switch back to weighing...

Страница 51: ...inebea Intec logo is displayed briefly after which the device is in normal weighing mode t Remove the preload weight from the weighing platform y The display shows the removed preload weight as a negative value t Press J to start Delete preload y The display CLR PREL appears briefly After a short period of time the preload will be deleted and the indicator will automatically switch back to weighin...

Страница 52: ...r adjustment p For printing Press briefly Prints GMP footer Press longer than 2 seconds K Toggles unit between normal and 10 fold higher resolution L Toggles the display between gross value net value plus tare and net value gross value minus tare c Quits an application or deletes an input character Press the key less than 2 seconds Deletes the entire input string Press the key longer than 2 second...

Страница 53: ...Return to initial state o Next higher menu level O Show items under selected entry Q Scroll up in the input output window q Scroll down in the input output window l Set the selected menu parameter Numeric Input via the Keypad t To enter numbers one digit at a time Press 0 1 2 9 t To save input press the corresponding key For example press to save manual tare input Text Input via the Keypad t Press...

Страница 54: ...ss the a key y ABC is displayed t Press the 1 key y The corresponding character selection is displayed The flashing cursor marks the first character t Press the F1 soft key l or wait 2 seconds y The special character appears on the display Deleting characters t Press c Deleting entire input string t Press the F4 soft key ESC Saving Input t Enter a value t Press the corresponding soft key e g 1 ID ...

Страница 55: ... weighing platform as a tare weight t Place the tare object on the weighing platform t Press y The value is applied as the tare value Input Through the Digital Control Port You can connect a remote hand switch or foot switch to the input control line for use with all application programs The following functions can be assigned in the Setup menu Device Parameters Control I O Ports Universal Switch ...

Страница 56: ...ent weighing range of the active weighing platform with multiple range scale connected Symbols and Their Meaning b The Busy symbol appears when the scale is processing a function activated by pressing a key The plus or minus sign of the weight or other measured value o Verified models only Identifies zero as a weight value after the scale or the active weighing platform has been zeroed Measured Va...

Страница 57: ...bar graph shows the percentage of the weighing platform s capacity that is used up by the load on the scale gross value 0 Lower load limit 100 Upper load limit The following symbols indicate tolerance levels for checkweighing Bar graph showing 10 intervals Minimum for checkweighing Target value for checkweighing Maximum for checkweighing Application Symbols A Symbol for the Counting application Y ...

Страница 58: ...Confirm parameter t M or oo Exit Setup menu Menu display Display Mode for Configuration and Information Setup Status line Input output window Soft key labels The display is divided into three sections Status Line Indicates the function of the current screen page In the Setup program this line shows the path to the data displayed Input and Output Window For input and display of detailed information...

Страница 59: ...selected Factory set parameters are identified by an o in the list starting on the next page Setting the Language You can choose from the following languages for displaying information Deutsch English factory setting U S mode Français Italiano Español Example Selecting U S Mode for the language t Press e to turn on the device t Press M y The menu appears on the display t Press the q soft key sever...

Страница 60: ... soft key several times to select the Device parameters line t Press the O soft key t Press the q soft key several times to select the Password line y The input line appears on the display t Enter numbers and or letters via the keypad see Numeric Input via the Keypad and Letter Input via the Keypad t Press the l soft key to save the setting t Press M or oo to exit the Setup menu 60 Operating Instr...

Страница 61: ...ric Input via the Keypad and Letter Input via the Keypad t Press the l soft key y The device selection appears on the display t Press the q soft key several times to select the Password line y The input line appears on the display t Press c several times to delete the password t Press the l soft key to save the delete If not yet saved the process can be canceled using the ESC soft key t Press M or...

Страница 62: ... BVers 01 63 02 Ser no 12345678 SETUP DEVICE WP 1 Internal WP 2 off COM1 Data communcations SBI Baud rate 1200 Baud Parity Odd Number of Stop bits 1 Stop bit Handshake mode Hardware 1 char Number of data bits 7 Bit Data output Printout printer 1 etc 62 Operating Instructions Combics Complete Scales Operating Design ...

Страница 63: ... manual adjust isoCAL function 3 o Off Adjustment prompt Activate external adjustment 2 o Activated Deactivated Parameter for external weight Cal adj weight Calibration adjustment unit Grams g Kilograms kg Tons t Pounds lb Adapt filter Min vibration o Normal vibration Strong vibration Extreme vibration Application filter o Final readout Filling mode Low filtering Without filtering Stability range ...

Страница 64: ...wt 1 Milligrams mg 1 Parts per pound lb 1 Chinese taels tlc 1 Mommes mom 1 Austrian carats K 1 Tola tol 1 Baht bat 1 Mesghal MS 1 Tons t lb oz Display accuracy 1 o All digits One less for wt change Index 1 1 Index 2 1 One less 1 Zero range 1 percent max cap o 2 percent max cap Zero at power on 2 percent max cap 5 percent max cap o 10 percent max cap Tare zero at power on o On Off Only zero at powe...

Страница 65: ... Milligrams mg 1 Parts per pound lb 1 Chinese taels tlc 1 Mommes mom 1 Austrian carats K 1 Tola tol 1 Baht bat 1 Mesghal MS 1 Tons t lb oz Display accuracy 2 o All digits One less for wt change Index 1 1 Index 2 1 One less 1 Factory settings weighing parameters only Yes o No o Factory settings 1 Not available on devices verified for use in legal metrology 2 Depends on weighing platform type Operat...

Страница 66: ...t change Index 1 1 Index 2 1 One less 1 Factory settings weighing parameters only Yes o No Off COM 1 when the WP is assigned to this interface COM 2 when the WP is assigned to this interface UNICOM when the WP is assigned to this interface Device parameters WP 2 RS 232 3 o ADC 232 similar to Internal menu for WP 1 RS 485 3 o ADC 485 similar to Internal menu for WP 1 o Off COM 1 similar to WP 1 COM...

Страница 67: ... o Hardware 1 char Number of data bits o 7 bits 8 bits Data output On request without stability o On request after stability Automatic without stability o 1 display update 2 display updates 10 display updates 100 display updates On request without stability o 1 display update 2 display updates 10 display updates 100 display updates Printout printer 1 Printout printer 2 Protocol printer 1 without s...

Страница 68: ...PI RS 232 SMA Baud rate 150 baud 300 baud 600 baud o 1200 baud 2400 baud 4800 baud 9600 baud 19200 baud Parity Space 1 Odd Even o None 2 Number of stop bits o 1 stop bit 2 stop bits Handshake mode Software handshake o Hardware 1 char Number of data bits o 8 bits o Factory settings 1 Not with 8 data bits 2 Not with 7 data bits 68 Operating Instructions Combics Complete Scales Operating Design ...

Страница 69: ...sal Baud rate 150 baud 300 baud 600 baud 1200 baud 2400 baud 4800 baud o 9600 baud 19200 baud Parity Odd Even o None Number of stop bits o 1 stop bit 2 stop bits Handshake mode o Software handshake Hardware 1 char Number of data bits o 8 bits o YDP04IS YPD05 o Strip Label Label manual form feed YDP21 Printer 2 1 Similar to printer 1 o Factory settings 1 Max 2 printers can be configured Operating I...

Страница 70: ...an be selected DeviceNet MAC ID 1 to 62 selectable Ethernet Printer 1 1 Similar to COM 1 Printer 2 1 Similar to COM 1 Analog output Analog output value Net value o Zero to max load Min Max values Gross value o Zero to max load Min Max values Analog Out Error Signal o High level 20 mA Low level 0 4 mA When menu is open or during calibration 0 4 mA on this interface Analog Out Trim o Min 0 4 mA Trim...

Страница 71: ...her resolution key L Net gross value key Level internal output ports o All active low All active high Bar code o Reference Tare ID1 Input Input without activating function External keyboard Config printout Headers Line 1 Line 2 ID codes ID1 ID2 ID3 ID4 ID5 ID6 ISO GLP GMP printout o Off For several application results Date time o Date with time Date only Once at stability o Off On FlexPrint o Off ...

Страница 72: ... dependent values Comp Printout after saving val 2 Similar to individual Total Printout after pressing CF 2 Similar to individual Printer 2 1 Similar to printer 1 Decimal separator o Period Comma Data records e g alibi memory product data memory All data records o Specified quantity of data records Printout factory setting Yes o No o Factory settings 1 Max 2 printers can be configured 2 Only when ...

Страница 73: ...y blocked CF key blocked Softkey 5 key blocked Softkey 4 key blocked Softkey 3 key blocked Softkey 2 key blocked Softkey 1 key blocked Application switch disabled Setup key blocked Mem key blocked Display Contrast 0 1 2 3 4 o 5 6 7 Backlighting o On Auto shutoff acc to timer Model o 2 1 Automatic shutdown Auto off acc to timer o No automatic shutoff Timer o 1 1 minute 2 2 minutes 5 5 minutes Main ...

Страница 74: ... 0 Int output 2 Equal 0 Int output 3 Heavier 0 Int output 4 Set 0 Read internal inputs Int input 1 0 Passwords Password SQmin Display o No Yes GMP print o No Yes Alibi memory Alibi memory period In days 90 o Factory settings 74 Operating Instructions Combics Complete Scales Operating Design ...

Страница 75: ...acc 1 Access switch 2 WP 2 see WP 1 3 WP 3 see WP 1 FlexInfo ID V Alibi memory Date Time Transaction no Language Deutsch o English U S mode Français Italiano Español o Factory settings 1 Depending on configuration prior to verification either latitude altitude or gravitational acceleration 2 Not displayed for internal ADC Operating Instructions Combics Complete Scales 75 Operating Design ...

Страница 76: ...ces Automatic printout Automatic data output see Data Interfaces Restore factory settings in the Setup menu Application parameters Factory settings only application Yes Soft Key Functions ID Enter up to six ID codes for identifying results on the printout 1st ID Save the value entered as the first ID code Preparation t Select Setup Press M t Select Application parameters Press the O soft key t Sel...

Страница 77: ...ere refer to the scale intervals for the connected weighing platform If the interval is 1 g and 1000 digits are required the minimum load is 1000 g 1000 intervals If the weighing platform interval is 5 g and the same number of digits as above are required the minimum load is 5000 g Once the minimum load has been exceeded the weighing platform is tared automatically and or an automatic printout is ...

Страница 78: ...er Operating Menu Setting Device parameters Bar code oReference value The value is applied and saved automatically Using a barcode scanner to enter an identifier Identifiers can be entered via a barcode scanner Operating Menu Setting Device parameters Bar code oID The value is applied and saved automatically Scanning barcodes directly You can directly scan a barcode using the barcode scanner Opera...

Страница 79: ... Counter is updated automatically when linearization calibration adjustment preload change is successful user calibration adjustment or linearization weight is changed When the following parameters are changed Function J key Zeroing range Initial zero point range The above parameters are reset to factory settings Configuration Counter Features Counter limited to 9999 Counter at P 0000 for hardware...

Страница 80: ...llowing settings are available All keys unblocked default setting All blocked except e and H Alphanumeric keys blocked n blocked blocked blocked k blocked J blocked p blocked K blocked L blocked c blocked Soft key 1 5 blocked D blocked H blocked c blocked Automatic Shutoff of Combics You can specify that the indicator will shut off after a set time has elapsed in the Setup menu under Device parame...

Страница 81: ... empty container on the scale y The container weight is displayed t Press the key to tare the scale Note If the automatic tare function is enabled you do not need to press the key The tare weight is saved automatically when you place the container on the platform t Wait until a zero value is displayed together with the NET symbol net weight t Place sample on the platform t Wait until the weight un...

Страница 82: ... the weighing platform at any time by pressing t Press the 25 key to enter a known tare weight via the keypad in this example 0 25 kg t Press the key to apply the tare weight t Place the container with sample on the platform t Read the results t Press the L key to toggle the display to the net weight value y The display shows the gross weight in this example 0 250 kg for the container plus 2 000 k...

Страница 83: ...dentifiers Start of GMP footer only if GMP compliant printout is configured End of GMP footer 24 02 2016 10 15 Model CAW3S Ser no 12345678 Vers H0 111 031115 BVers 01 63 02 Ser no C 12345678 EISENSCHMIDT GOETTINGEN Chargen Nr 123456 Kunde SCHULZE 24 02 2016 10 15 G 2 250 kg T 0 000 kg PT2 0 250 kg N 2 000 kg 24 02 2016 10 16 Name Operating Instructions Combics Complete Scales 83 Operation ...

Страница 84: ... the key to tare the scale Note If the automatic tare function is enabled you do not need to press the key The tare weight is saved automatically when you place the container on the platform t Wait until a zero value is displayed together with the NET symbol net weight t Place a packaged sample in the container t Press the 25 key to enter the known tare weight of the package via the keypad in this...

Страница 85: ...ess the key to apply the entered value y The tare values are deleted The gross value appears on the display t Press the p key to print the results G 6 433 kg T 4 183 kg PT2 0 250 kg N 2 000 kg G 6 433 kg T 0 000 kg N 6 433 kg Operating Instructions Combics Complete Scales 85 Operation ...

Страница 86: ... legal metrology the scale should be adjusted via the internal adjustment equipment at the installation location see chapter Configuring Weighing Platforms section Internal Adjustment 3 The temperature range C listed on the ID label should not be exceeded during operation For Servicing External adjustment for verified scales of accuracy class l External adjustment is blocked in legal metrology swi...

Страница 87: ...alibrate first then adjust automatically or manually not for verified weighing instruments under Calibration adjustment Cal adj sequence For Minebea Intec IS weighing platforms only Flashing WP symbol as adjustment prompt if more than one weighing platform is connected the platform number is also displayed under Calibration adjustment CAL key function Block Unblock external adjustment under Calibr...

Страница 88: ...xt cal adj user defined wt Key blocked Cal adj sequence Cal then auto adj o Cal then manual adj isoCAL function for Minebea Intec IS weighing platforms only o Off Adjustment prompt Activate ext adj o Activated Deactivated Parameter for external weight o Factory setting t To save settings and exit the Setup menu Press the M key or the oo soft key Procedure t Zeroing the scale by pressing the key t ...

Страница 89: ...stopped by pressing the key t Start adjustment via the J key cancel calibration adjustment via y The adjustment weight is displayed once adjustment is finished y A GMP compliant printout is generated Ext Calibration Nom 10000 g Diff 1 g 24 02 2016 10 15 Model CAW3S Ser no 12345678 Vers H0 111 031115 BVers 01 63 02 Ser no C 12345678 Ext Calibration Nom 10000 g Diff 1 g Ext adjustment Diff 0 g 24 02...

Страница 90: ...load and zero the scale t Press J to start the internal calibration C INTER C EXT DEF C Inter y The display C INTER appears while the internal adjustment weight is placed on the platform CI C EXT DEF Just Intern y Then the display CI appears y After a brief pause the difference since the last span adjustment is displayed t Press J to start internal adjustment or press to cancel the calibration adj...

Страница 91: ...e technician will determine the permitted minimum sample quantity and set this on your balance using your QA system s guidelines He will document this setting via a Weighing module test as per USP certificate in which the measurements and min sample quantity are logged The SQmin function ensures that the weighing results correspond to USP guidelines These SQmin settings cannot be changed by the us...

Страница 92: ...nd save using the l soft key or cancel using ESC t Then enter the values for WP 2 and WP 3 t Press M or oo to exit the Setup menu t Slide the menu access switch to the right closed position and reattach the cap y The device is now in normal weighing mode Configuration in Weighing Mode t In the Device parameters menu select SQmin and open using the O soft key t Select Display t Open using the O sof...

Страница 93: ... the p key to generate the printout t Place another sample on the scale y The minimum sample quantity is exceeded no symbol k t Press the p key to generate the printout t Briefly press the k key to toggle between the measured value and SQmin value y The value for the minimum sample quantity is displayed for four seconds Operating Instructions Combics Complete Scales 93 Operation ...

Страница 94: ...haracters are printed Enter up to 21 characters for the value of the ID code Press the ID soft key to activate the input mode Enter the first ID value directly through the numeric keypad Press the 1st ID soft key to save the value Individual characters of the ID can be deleted using the key Complete ID code values can be deleted using the Delete soft key If both the name and value fields are empty...

Страница 95: ...pad see also Operating Design Text Input via the Keypad t Press the l soft key to save the entry t Press M or oo to exit the Setup menu Example Enter ID code values The values 123456 and Smith should be entered for ID 1 and ID 2 t Press the ID soft key y The first line is selected t Press the key to enter 123456 via the keypad see also Operating Design Number Input via the Keypad t Press the l sof...

Страница 96: ...canner or a keyboard can be connected via the PS 2 socket or the corresponding screw terminals IP69K Features IP44 indicator Connect via a 25 pin D Sub female connector IP69K indicator Route the connecting cable from the peripheral device to the indicator via a cable gland The free cable ends are connected via the screw terminals 3 Warning when using third party RS 232 connecting cables The pin as...

Страница 97: ...dress3 0 1 2 31 SBI Manual data output Withoutstability afterstability configurableprintout SBI Auto data output Without stability at stability SBI Output format 16 or 22 characters Printout of application data Output of a configurable printout Analog UniCOM interface optional Level 4 20 mA 0 20 mA 0 10 V Power supply Internal Factory setting 4 20 mA internal power supply Connection CAISL3 devices...

Страница 98: ...r External keyboard The following devices can be connected to the optional UniCOM universal interface WP3 RS 232 RS 485 PC RS 232 interface Second printer external power source required Remote display Digital I O galvanically isolated Current interface 0 4 20 mA voltage 0 10 V PLC with Profibus DP or DeviceNet Ethernet Connection Options for other Weighing Platforms The Combics 3 model enables you...

Страница 99: ...YD21 Label man form feed COM2 oOff WP 3 RS 232 SBI standard version 9600 baud SBI trade version for legal metrology oIS 232 ADC 232 Data oSBI communication XBPI 232 SMA Printer 1 1 YDP20 or YDP14IS Strip Printer 2 1 o Label Universal printer YDP05 oStrip Label YDP21 Label man form feed UNICOM oOff option WP 3 RS232 RS485 Data oSBI communication XBPI 232 XBPI 485 SMA Profibus DeviceNet Ethernet Pri...

Страница 100: ...t value Only 14 characters are available if you disable the header This is carried out in the Line format menu item see also Setup Overview in Configuration Automatic Data Output SBI You can have the weight readout printed automatically1 This printout can be generated after a certain number of display updates You can also configure whether or not the auto print function is dependent on the stabili...

Страница 101: ... is active The transaction number is returned Only effective with setting Protocol printout without stability 6 1 8 a4xx_ Alibi memory Read content according to the transaction number specified under xx kT_ Trigger key tare kNW_ Trigger n key function toggle the weighing platform kZE_ Trigger key function zero the instrument kCF_ Trigger c key function x1_ Output model designation of active weighi...

Страница 102: ... 4 5 6 7 8 9 10 11 12 13 14 15 16 A A A A A A A A E E E CR LF or A A A A A A A A E E E CR LF or CR LF Signs the sign can be hidden under Device parameters COMx Data communications SBI Sign format Space A Digit or letter max 7 characters plus decimal point E Unit symbol1 1 to 3 letters followed by 2 0 spaces CR Carriage return LF Line feed Special Codes Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 ...

Страница 103: ...9 10 11 12 13 14 15 16 17 18 19 20 21 22 KK K K K K A A A A A A A A A E E E CR LF KK K K K K A A A A A A A A A E E E CR LF CR LF K ID code character right justified with spaces Plus or minus sign Space A Digit or letter max 7 characters plus decimal point E Measurement unit symbol1 1 to 3 letters followed by 2 0 spaces CR Carriage return LF Line feed Special Codes 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...

Страница 104: ... from Counting piece count and Neutral Measurement applications mDef Target value for animal weighing x Net Animal weighing results Setp Target value for checkweighing Diff W Absolute difference e g in kg in Checkweighing Lim Deviation in in Checkweighing Max Upper tolerance for checkw Min Min tolerance for checkw Stat Status Classx Classification class x Limx Class limit D Percentage as loss Prc ...

Страница 105: ...nit symbol is not included with output External Keyboard Functions PC Keyboard For settings go to the Setup menu under Device parameters Bar code External keyboard ThealphanumerickeycodesimplementedherearespecifictotheGermankeyboard layout The following alphanumeric characters are used some require the Shift key a z A Z 0 9 space and these characters _ Function keys PC keyboard Combics 3 F1 key F2...

Страница 106: ...s o 1 printout 2 printouts Single o Max 30 print items printout 1 can be selected o ID1 ID4 Component o Max 30 print items Printout 1 can be selected Total data Max 30 print items printout can be selected Printer 2 as for Printer 1 Factory setting Printout Reset Do not reset o Factory setting 1 Multiple selections possible 2 When the minimum load is exceeded can be set under menu item Application ...

Страница 107: ...ks that can be activated or deactivated via multiple selection in the menu For the Totalizing and Net total Formulation applications the totalizing results printout can be configured independent of the individual component printout The individual information blocks are shown below with detailed explanations Samples of complete printouts are provided following the end of this section Headers 2 head...

Страница 108: ...off factory setting GMP compliant printout for multiple application results The GMP header is included from the first printout generated subsequent to the activation of the GMP printout The GMP footer is printed after several measurement results by pressing and holding the p key e g for component printing menu item For multiple application results In this case the T symbol remains visible until th...

Страница 109: ...ER LINE1 HEADER LINE2 14 01 2016 09 43 Ser no 80705337 G 1 402 kg T 0 200 kg N 1 202 kg Display with ID and weighing platform Ser no 80705337 G 1 402 kg T 0 200 kg N 1 202 kg Counting Application The initialization data block contains the reference sample quantity and the reference sample weight nRef 10 pcs wRef 0 035 kg G 1 402 kg The results block contains gross net and tare weight T 0 212 kg N ...

Страница 110: ...k contains the reference percentage and reference weight Percentage residual pRef 100 Wxx 2 100 kg G 1 859 kg The results block contains gross net and tare weights as well as the percentage T 0 200 kg which is shown as either the loss or the residual amount N 1 659 kg Prc 79 and the piece count as a result Percentage loss pRef 100 Wxx 2 100 kg G 0 641 kg The results block contains gross net and ta...

Страница 111: ...Relation to target value In the OK range the deviation from the target weight is printed as a percentage and absolute deviation In the nonconforming range HH is printed for exceeding the weight and LL for falling below the weight OK range in the weight and tolerance limit display Nonconforming range in the weight display Setp 1 300 kg Min 1 235 kg Max 1 365 kg G 1 312 kg T 0 000 kg N 1 312 kg Lim ...

Страница 112: ...2 Complete standard printout configuration is printed for each transaction 14 01 2016 09 43 G 2 400 kg T 0 200 kg N 2 200 kg n 2 G 2 400 kg Standardausdruck T 0 200 kg The transaction counter is not printed N 2 200 kg Print menu parameters MENU All active sub items of the currently displayed menu are printed SETUP WP1 1 1 1 1 1 2 1 2 1 1 3 2 1 18 1 18 1 CAL 10 000 kg etc 112 Operating Instructions...

Страница 113: ...n adjustment printout 24 02 2016 10 15 Model CAW3S Ser no 12345678 Vers H0 111 031115 BVers 01 63 02 Ser no C 12345679 External calibration Targ 30 00 kg Diff 0 03 kg External adjustment Diff 0 00 kg 24 02 2016 10 15 Name Setting the preload printout 24 02 2016 10 15 Model CAW3S Ser no 12345678 Vers H0 111 031115 BVers 01 63 02 Ser no C 12345679 Set preload completed 24 02 2016 10 15 Name Operatin...

Страница 114: ...ng printout with multiple results example 2 results 24 02 2016 10 15 Model CAS3E Ser no 12345678 Vers H0 111 031115 BVers 01 63 02 Ser no C 12345679 HEADER LINE1 HEADER LINE2 4 02 2016 10 15 G 2 40 kg T 0 20 kg N 2 20 kg HEADER LINE1 HEADER LINE2 4 02 2016 10 15 G 3 40 kg T 0 30 kg N 3 10 kg 4 02 2016 10 15 Name 114 Operating Instructions Combics Complete Scales Data Interfaces ...

Страница 115: ...rs are displayed for the duration of the error via the ERR error code Temporary errors are displayed for 2 seconds via the INF error code then the program automatically switches back to normal weighing operation Operating Instructions Combics Complete Scales 115 Error Codes ...

Страница 116: ... Function not allowed Function not allowed in scales verified for Contact your local Service Center for for verifiable weighing use in legal metrology information on changing settings INF 08 The load on the scale is too heavy to zero Check whether Tare zero at power on was complied the readout with in your configuration INF 09 Taring is not possible when the scale gross Zero the scale weight is ze...

Страница 117: ...ebeginningtoopentheequipment Properfitting of all surfaces is essential for the IP rating of the housing for this reason the device must be opened and closed by a certified technician Cleaning 2 Disconnect the power supply to the scale before cleaning unplug the power cord from the mains supply Cleaning the interior of the weighing platform CAW models only t Remove the load plate Special care is r...

Страница 118: ...lay and control unit and press it over the edge of the front and rear side of the device until it is fixed in place Safety Inspection Safe operation of the device is no longer ensured when The device or the mains connecting lead shows visible damage The integrated power supply for the indicator no longer functions properly The device has been stored for a relatively long period under unfavorable c...

Страница 119: ...rators For disposal in Germany and in the other member nations of the European Economic Area EEA please contact our local service technicians or our Service Center in Bovenden Germany Minebea Intec Bovenden GmbH Co KG Leinetal 2 37120 Bovenden WEEE Reg Nr DE58091735 In countries that are not members of the European Economic Area EEA or where no Minebea Intec subsidiaries or dealerships are located...

Страница 120: ...interval mode or 2x 3000e Multiple range mode according to EN 45501 Digital protective interface According to EN45501 Data interface Bidirectional RS 232C with control outputs 5 V TTL standard built in as standard Model specific Data Model type CAW P L CAW P I CAW P S CAW P BCE CAW P NCE 2 x 3000e Readability Weighing range 1 Weighing range 2 Weighing 15000d 30000d 60000d 1x3000e Maximum Readabili...

Страница 121: ...15 5 15 5 30 10 6 2 15 5 30 10 30 10 60 20 15 5 30 10 60 20 60 10 60 20 150 50 30 10 60 20 150 50 120 20 150 50 300 100 60 20 150 50 300 100 300 50 300 100 600 200 150 50 300 100 600 200 600 100 600 200 1500 500 300 100 600 200 1500 600 1200 200 1500 500 3000 1000 600 200 1500 500 3000 1000 3000 500 TheweighingrangespermittedforuseinlegalmetrologyarelistedintheDeclarationofConformity Therespective...

Страница 122: ...800 600 117 300IG H 300 2 800 600 117 300IG LCE 300 20 20 1000 800 600 117 Only kg as weight unit Type designation Combics high resolution complete scales CAS3E Application Base frame design of platform Weighing range Platform dimensions Resolution Resolutionverifiableversions CAS3 E 16 ED H HCE E 34 ED H HCE E 64 ED S SCE Order code E Base frame epoxy resin coated CAS3G Application Base frame des...

Страница 123: ...E SPC 06HCE CASLE SPC 3HCE CASLE SPC 3ICE CASLE SPC 6HCE CASLE SPC 6SCE Weighing capacity kg 0 62 3 1 3 1 6 1 6 1 Readability g 0 001 0 01 0 1 0 01 0 1 verification scale intervalt e g 0 01 0 1 0 1 0 1 0 1 Min load 0 02 0 5 5 0 5 5 Accuracy class K K K K K Resolution code HCE HCE HCE HCE SCE Calibration weight value in grams 500 2 000 2 000 5 000 5 000 Accuracy class E2 E2 E2 E2 F2 Preload kg 0 0 ...

Страница 124: ...124 Operating Instructions Combics Complete Scales Specifications Device Dimensions Dimensions of CAS platforms Scale Drawings All dimensions are given in millimeters CAS E ...

Страница 125: ...Operating Instructions Combics Complete Scales 125 Specifications CAS G FE ...

Страница 126: ...126 Operating Instructions Combics Complete Scales Specifications ...

Страница 127: ...Operating Instructions Combics Complete Scales 127 Specifications CAS G IG ...

Страница 128: ...128 Operating Instructions Combics Complete Scales Specifications ...

Страница 129: ...Operating Instructions Combics Complete Scales 129 Specifications Dimensions of indicator Scale Drawings All dimensions are given in millimeters ...

Страница 130: ...0 Operating Instructions Combics Complete Scales Specifications CASE SPC 06HCE All dimensions are given in millimeters 165 8 4 9 2 5 1 4 5 5 1 5 4 9 8 1 0 6 5 206 236 9 8 1 0 6 5 8 M x1 237 14 266 152 35 ...

Страница 131: ...SLE SPC 6HCE All dimensions are given in millimeters 1 8 2 1 9 4 8 4 9 2 5 1 1 4 5 1 2 3 1 1 6 5 I P 6 5 1 2 5 2 3 6 2 0 6 1 8 2 1 9 4 8 M x 1 3 5 1 4 2 6 6 2 3 7 1 5 2 1 8 2 1 9 4 8 4 9 2 5 1 1 4 5 1 2 3 1 1 6 5 I P 6 5 1 2 5 2 3 6 2 0 6 1 8 2 1 9 4 8 M x 1 3 5 1 4 2 6 6 2 3 7 1 5 2 ...

Страница 132: ...132 Operating Instructions Combics Complete Scales Specifications CASE SPC 16HCE CASLE SPC 35HCE All dimensions are given in millimeters 350 1 3 3 5 1 4 1 5 8 M x1 304 35 192 240 257 ...

Страница 133: ...CASS SPC 16HCE CASS SPC 35HCE All dimensions are given in millimeters Operating Instructions Combics Complete Scales 133 Accessories ...

Страница 134: ...34 Operating Instructions Combics Complete Scales Specifications ...

Страница 135: ...1 01CISLM3 Adapter cable for CAIS indicator YCC02 D09M6 3 paper rolls for YDP05 and ZD420d 60 mm 75 m thermal paper 69Y03090 Labels small 58 mm 30 mm 1000 labels 69Y03092 Labels medium 58 mm 76 mm 500 labels 69Y03093 Labels large 58 mm 100 mm 350 labels 69Y03094 UniCOM interface CAW1P CASLE SPC CAW1 2 3S CAW3P CAS3 CASE SPC Installation options For installation For installation For installation in...

Страница 136: ...printer YDP05 14IS 9 pin D SUB plug 6 m YCC02 D09M6 for printer YDP21 or PC 9 pin D SUB socket 6 m YCC02 D09F6 for scales 25 pin D SUB plug 6 m YCC02 D25M6 for various accessories 25 pin D SUB socket 6 m YCC02 D25F6 for scales 12 pin round plug 6 m YCC02 R12M6 for various accessories and IS platforms 12 pin round plug socket 6 m YCC02 R12F6 open cable ends 6 m YCC02 RELAIS02 Ethernet connection ca...

Страница 137: ...ification in Germany Scale verification for legal metrology is valid until the end of the calendar year after next If the scale is used for fill level control in accordance with legislation on prepackaging verification is valid until the end of the following calendar year Re verification must currently be carried out by a weights and measures official Re verification should be requested in good ti...

Страница 138: ...138 Operating Instructions Combics Complete Scales Declarations of Conformity ...

Страница 139: ...Operating Instructions Combics Complete Scales 139 Declarations of Conformity ...

Страница 140: ...40 Operating Instructions Combics Complete Scales Declarations of Conformity 1 5 ...

Страница 141: ...ón de armonización pertinente de la Unión Europea 6 Referencias a las normas armonizadas pertinentes utilizadas o referencias a las otras especificaciones técnicas respecto a las cuales se declara la conformidad 3 5 eesti keel et français fr hrvatski hr Vastavusdeklaratsioon Déclaration de conformité Izjava o sukladnosti 1 Tootemudel tootenumber kehtib vaid järgmise projekti puhul 2 Tootja nimi ja...

Страница 142: ...eclaração acima descrito s está ão em conformidade com a legislação aplicável de harmonização da União 6 Referências às normas harmonizadas aplicáveis utilizadas ou às outras especificações técnicas em relação às quais é declarada a conformidade 1 Modelul de produs Număr produs valabil numai pentru numărul proiectului 2 Denumirea și adresa producătorului 2 1 și a reprezentantului său autorizat 2 2...

Страница 143: ...Operating Instructions Combics Complete Scales 143 EU type examination certificate ...

Страница 144: ...144 Operating Instructions Combics Complete Scales EU type examination certificate ...

Страница 145: ...Operating Instructions Combics Complete Scales 145 Test Certificate ...

Страница 146: ...146 Operating Instructions Combics Complete Scales 146 Evaluation Certificate ...

Страница 147: ...Operating Instructions Combics Complete Scales 147 14 Evaluation Certificate ...

Страница 148: ...148 Operating Instructions Combics Complete Scales Evaluation Certificate ...

Страница 149: ... Operating Instructions Combics Complete Scales 149 Certificate ...

Страница 150: ...150 Operating Instructions Combics Complete Scales Plates and Markings Minecomb Type of weighing instrument Minecomb Type of indicator TA EC type approval certificate T11379 test certificate D09 11 02 ...

Страница 151: ...eight instrument already verified K Plates and Markings CAW3S1 60FE NCE CAW3S1 60FE NCE CAAPS1 60FE NCE Type of weighing instrument Minecomb Type of indicator TA EC type approval certificate T11379 test certificate D09 11 02 Type of weighing instrument Minecomb Type of indicator TA EC type approval certificate T11379 test certificate D09 11 02 Mark for EC verification metrology sticker ...

Страница 152: ...ng Instructions Combics Complete Scales Plates and Markings CAW3S1 60FE NCE CAW3S1 60FE NCE Type of weighing instrument Minecomb Type of indicator TA EC type approval certificate T11379 test certificate D09 11 02 ...

Страница 153: ...Operating Instructions Combics Complete Scales 153 ...

Страница 154: ...154 Operating Instructions Combics Complete Scales Plates and Markings ...

Страница 155: ...Operating Instructions Combics Complete Scales 155 Plates and Markings ...

Страница 156: ...156 Operating Instructions Combics Complete Scales Plates and Markings ...

Страница 157: ...Datenspeicher Protective mark self adhesive mark or seal only in case of existent approved data storage device S Typenschild Type plate T Sicherungsstempel selbstklebende Marke oder Plombe Protective mark self adhesive mark or seal S MD T MD S S S S Alibispeicher optional Approved data storage device optional frei unlocked Programmverriegelungs schalter Menu access switch gesperrt locked S Schnitt...

Страница 158: ...ification metrology M M Sicherungsstempel selbstklebende Marke oder Plombe Protective mark self adhesive mark or seal S Typenschild Type plate T Schilder und Marken CASE SPC Typ Minecowat Plates and Markings T S S Schnittstelle Terminal Inklusive Spannungsversorgung Interface terminal Inclusive power supply S M K S S gesperrt locked frei unlocked Programm verriegelungs schalter Menu access switch ...

Страница 159: ...nzeichnungsschild mit CE Zeichen der bereits geeichten Waage Descriptive plate with CE sign on a weighing instrument already verified M Beispiel für Schild mit metrologischen Daten Example of label with metrological data T Typenschild Wägemodul Type plate weighing module T MD Schilder und Marken CASE SPC Typ Minecowat Plates and Markings ...

Страница 160: ...i vorhandenem eichfähigem Datenspeicher Protective mark self adhesive mark or seal only in case of existent approved data storage device S Typenschild Type plate T Sicherungsstempel selbstklebende Marke oder Plombe Protective mark self adhesive mark or seal S MD MD T S S S S Alibispeicher optional Approved data storage device optional zum Wägemodul to the weighing module S frei unlocked gesperrt l...

Страница 161: ...mark self adhesive mark or seal S Schilder und Marken CASLE SPC Typ Minecowat Paltes and Markings zum Wägemodul to the weighing module S frei unlocked gesperrt locked Programmverriegelungs schalter Menu access switch zum Wägemodul to the weighing module frei unlocked S gesperrt locked Programmverriegelungs schalter Menu access switch ...

Страница 162: ...en für die EG Eichung Messtechnik M Mark for EC verification metrology sticker M Sicherungsstempel selbstklebende Marke oder Plombe Protective mark self adhesive mark or seal S Typenschild Type plate T Schilder und Marken CASLE SPC Typ Minecowat Paltes and Markings T S S Programm verriegelungs schalter Menu access switch frei unlocked gesperrt locked S K M Schnittstelle Terminal Inklusive Spannung...

Страница 163: ...zeichnungsschild mit CE Zeichen der bereits geeichten Waage Descriptive plate with CE sign on a weighing instrument already verified M Beispiel für Schild mit metrologischen Daten Example of label with metrological data T Typenschild Wägemodul Type plate weighing module T MD Schilder und Marken CASLE SPC Typ Minecowat Paltes and Markings ...

Страница 164: ...ly in case of existent approved data storage device S Typenschild Type plate T Sicherungsstempel selbstklebende Marke oder Plombe Protective mark self adhesive mark or seal S MD Schilder und Marken CASS SPC Typ Minecowat Plates and Markings T MD S S S S Alibispeicher optional Approved data storage device optional frei unlocked Programmverriegelungs schalter Menu access switch gesperrt locked S Sch...

Страница 165: ...chen für die EG Eichung Messtechnik M Mark for EC verification metrology sticker Kennzeichnungsschild mit CE Zeichen Descriptive plate with CE sign Typenschild Type plate T Schilder und Marken CASS SPC Typ Minecowat Plates and Markings T S S S S K M S gesperrt locked frei unlocked Programm verriegelungs schalter Menu access switch Schnittstelle Terminal Inklusive Spannungsversorgung Interface term...

Страница 166: ...ichung Messtechnik M Mark for EC verification metrology sticker M Sicherungsstempel selbstklebende Marke oder Plombe Protective mark self adhesive mark or seal S Typenschild Type plate T Schilder und Marken CASS SPC Typ Minecowat Plates and Markings T S S gesperrt locked Programm verriegelungs schalter Menu access switch Schnittstelle Terminal Inklusive Spannungsversorgung Interface terminal Inclu...

Страница 167: ...schild mit CE Zeichen der bereits geeichten Waage Example of descriptive plate with CE sign on a weighing instrument already verified M Beispiel für Schild mit metrologischen Daten Example of label with metrological data T Beispiel Typenschild Wägemodul Example of type plate weighing module T MD Schilder und Marken CASS SPC Typ Minecowat Plates and Markings ...

Страница 168: ...168 Operating Instructions Combics Complete Scales Plates and Markings ...

Страница 169: ...Operating Instructions Combics Complete Scales 169 Plates and Markings ...

Страница 170: ...170 Operating Instructions Combics Complete Scales Plates and Markings ...

Страница 171: ...Operating Instructions Combics Complete Scales 171 Plates and Markings ...

Страница 172: ...172 Operating Instructions Combics Complete Scales ...

Страница 173: ...Operating Instructions Combics Complete Scales 173 Certificate ...

Страница 174: ...174 Operating Instructions Combics Complete Scales ...

Страница 175: ...thout notice and does not represent a commitment on the part of Minebea Intec unless legally prescibed 1 Power connection Explosion risk area zone 2 or zone 22 Explosion safe area Indicator CAIS or complete scale CAW each with option Y2 2 Data transfer PC printer PLC etc For data transfer only and if required supply for the connected equipment e g printer there is no supply voltage going from the ...

Страница 176: ...on Y2 series or other weighing platforms suitable for zone 2 or 22 category 3G or 3D which may be supplied with at least 9 VDC 4 5 VDC ADC output with bridge rated voltage 9 VDC 4 5 VDC 3 Connecting weighing platforms 4 Connection to external equipment Potential equalization Potential equalization Indicator CAIS with option Y2 Analog weighing platforms load cells of the CAAP option Y2 series Conne...

Страница 177: ... jolts or vibrations and install so that heat can be sufficiently dissipated on all sides and external heat sources are located far enough away 7 The cable glands for the data cables must be tightened with a torque of 5 Nm The cable gland for the network cable must be tightened with a torque of 2 5 Nm cap and 1 5 Nm body Attach external connection cables securely to avoid damage and tensile loadin...

Страница 178: ...bsequently make sure that the connections are not corroded The grounding conductor of a power cord must be the same gauge as the current carrying wires N and L 19 All external cables even cables between load cells and junction box are only suitable for fixed installation and must be laid fixed Otherwise use cable glands designed according to EN 60079 0 and rounded at an angle of 75 minimum and a r...

Страница 179: ... opens with the Excel program MS Excel must already be installed on your computer A dialog box for selecting macros opens t Click on the Activate macros button y Note This window might not open depending on the settings in your computer system t All fields on the Data page highlighted in yellow must be filled out by a qualified person y A filled out sample document is available in the Documents fo...

Страница 180: ...180 Operating Instructions Combics Complete Scales Certificate US C ...

Страница 181: ...Operating Instructions Combics Complete Scales 181 Certificate US C ...

Страница 182: ...182 Operating Instructions Combics Complete Scales ...

Страница 183: ...e Access window appears on the display t Enter the general password see below via the keypad see Numeric Input via the Keypad t Press the l soft key y The device selection appears on the display t Press the q several times to select the Password line t Press the O soft key y The input line appears on the display t Read the old password or enter a new password max 8 characters t Press or c several ...

Страница 184: ...184 Operating Instructions Combics Complete Scales ...

Страница 185: ...Operating Instructions Combics Complete Scales 185 ...

Страница 186: ...186 Operating Instructions Combics Complete Scales ...

Страница 187: ...Operating Instructions Combics Complete Scales 187 ...

Страница 188: ...without prior written permission from Minebea Intec All rights reserved by Minebea Intec in accordance with copyright law The information and figures contained in these instructions correspond to the version date specified below Minebea Intec reserves the right to make changes to the technology features specifications and design of the equipment without notice Date March 2021 Printed in Germany on...

Отзывы: