4-7
CAUTION
z
It is necessary to wait at least 5 seconds after setting the control dial to the 0 position
before opening the vaporizer. This allows the pressure to balance and prevents fresh
gas and anesthetic agent vapor froming escaping from the vaporizer.
2. Select the correct filling adapter and anesthetic agent bottle. Screw the filling adapter firmly into
the anesthetic agent bottle. Before use, check that the color marks and names/symbols of anesthetic
agent on the filling adapter, anesthetic agent bottle and vaporizer must correspond to the anesthetic
agent used.
WARNING
z
Do not use a damaged filling adapter or an anesthetic agent bottle without collar. If a
bottle without collar is used, specific filling adapter cannot be identified, underlying the
risk of filling a wrong anesthetic agent.
CAUTION
z
If the connection between the filling adapter and anesthetic agent bottle is not
leak-tight, anesthetic agent may escape.
NOTE
z
If a new anesthetic agent bottle is partly empty, there may be a leak.
Filling adapter
Anesthetic agent bottle
Bottle neck collar
Содержание V60
Страница 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Страница 2: ......
Страница 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Страница 10: ...4 FOR YOUR NOTES ...
Страница 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Страница 48: ...5 4 FOR YOUR NOTES ...
Страница 56: ...6 8 FOR YOUR NOTES ...
Страница 66: ...9 4 FOR YOUR NOTES ...
Страница 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Страница 80: ...11 12 ...
Страница 84: ...A 4 FOR YOUR NOTES ...
Страница 86: ...B 2 FOR YOUR NOTES ...
Страница 89: ......
Страница 90: ...P N 046 001002 00 8 0 ...