4-12
5. Remove the filler cap and insert the bottle equipped with adapter into the filler receptacle.
Rotate the bottle gently to align the bottle filler adapter with the slots in the filler receptacle.
6. Press the bottle until the liquid begins to flow into the vaporizer.
CAUTION
z
If the connection between the filling adapter and anesthetic agent bottle is not
leak-tight, anesthetic agent may escape.
7. Check the filling level in sight glass during filling. When the maximum mark is reached, stop
pressing the bottle. And pull out the bottle slowly. If the vaporizer is filled above the
maximum mark by a few millimeters, the anesthetic agent will start to overflow through the
overflow hole.