Chapter 2 Principles
20
2.4
System Parameter
2.4.1 General
For the PM-7000, signals are collected by modules, and the results are transferred to the main
board through the adapter board, thus to process and display the data and waveforms.
Commands from the main board, as well as the status information of modules, are transferred
through the adapter board. In addition, the adapter board adapts and changes the power
supply. The structure of the whole system is shown in the following figure.
---
As shown in Figure 2-14, the five modules and measurement cables monitor and measure
NIBP, SpO
2
, ECG/RESP/TEMP, IBP/CO and CO
2
in real time, and send the results to the main
board for processing and displaying. If necessary, the results are sent to the recorder for
printing.
The parameter monitoring functions are described respectively in the following sections.
2.4.2 ECG/RESP
■
ECG
The PM-7000 has the following ECG functions:
1) Lead type: 3-lead, 5-lead, 12-lead
2) Lead way:
3-lead (1 channel):
5-lead (2 channels):
I, II, III
I, II, III, aVR, aVL, aVF, V