Chapter 5 Test and Material List
65
5.5.2 AG CALIBRATE
(
Agent>1.5%, CO2>1.5%, N2O>40%, O2>40%
)
1
、
Press ‘START CAL’, then input password ‘MINDRAY’ entering ‘CALIBRATE’ menu.
Note: Make sure the AG in ‘Measure’ mode not ‘Standby’ mode before you do calibrate.
Figure 5-10
Figure 5-11
2
、
Input each gas concentration value according to the label on the AG bottle label.
Note: If your monitor do not have O2 module, input ‘0.0’ in O2 item.
3
、
Open AG cover, wait for the display value stabilization.
4
、
If the display value does not accord with the input value, please press ‘CALIBRATE’ item
and exit.
AG concentration must fit the following requirements:
Agent>1.5%, CO2>1.5%, N2O>40%, O2>40%
Содержание PM-7000
Страница 1: ...PM 7000 Patient Monitor Service Manual ...
Страница 2: ......
Страница 58: ...Chapter 3 Product Specifications 46 line 2 5m ENF 350 ms Delay time 4s ...
Страница 81: ......
Страница 82: ...P N 7001 20 67407 1 0 ...