Chapter 3 Product Specifications
45
5 to 10
±
0.4
10 to 15
±
0.6
15 to 18
±
1
>18 Not
specified
0 to 1
±
0.15
1 to 5
±
0.2
5 to 8
±
0.4
Sev
> 8
Not specified
0 to 1
±
0.15
1 to 5
±
0.2
Enf, Iso, Hal
> 5
Not specified
0 to 25
±
1
25 to 80
±
2
O2 (Optional)
80 to 100
±
3
2 to 60 BrPM
±
1 BrPM
AwRR
> 60 BrPM
Not specified
Alarm range
CO2:
AwRR:
0 to 10 % (0 - 76 mmHg)
2 to 100 BrPM
Apnea alarm delay
AwRR
:
20 - 40 s
Refreshing rate
1s
Calibration
Yearly calibration requested.
Calibration stability
After module being used for 12 consective months, the error is <
1%
CO
2
250 ms (fall time 200 ms)
N
2
O
250 ms
O
2
600ms
HAL, ISO, SEV, DES
300 ms
Rise time (10 % to 90 %)
Sampling flow 120ml/min,
using the DRYLINE™
water trap and neonatal
DRYLINE™ sampling
line (2.5m)
ENF
350 ms
CO
2
250 ms (fall time 200 ms)
N
2
O
250 ms
O
2
500ms
Rise time (10 % to 90 %)
Sampling flow 200ml/min,
using the DRYLINE™
water trap and adult
DRYLINE™ sampling
HAL, ISO, SEV, DES
300 ms
Содержание PM-7000